EM TEST UCS 500N5 Transient Generator
Description
UCS 500N5 Multifunctional Test Generator for Transients (EFT/Burst, Surge & Power Fail) up to 5.5 kV as per IEC/EN 61000-4-4, 61000-4-5 and 61000-4-11
EM TEST UCS 500N5 Transient Generator is the most versatile tester to cover transient and power fail requirements according to international standards (basic and generic standards) and product/product family standards. The UCS 500N5 is the most economical solution for tests during development as well as for full-compliant immunity tests and CE Marking for single phase DUT with the ability to be extended for testing three-phase DUTs by means of an automatically controlled external coupling network up to 100A.
MODEL OVERVIEW
UCS 500N5E Electrical Fast Transient/Burst Generator up to 5.5 kV as per IEC/EN 61000-4-4
UCS 500N5P Power Fail Simulator for Dips & Short Interruptions as per IEC/EN 61000-4-11 and 61000-4-29
UCS 500N5V Combination Wave (Surge only) Generator up to 5kV as per IEC/EN 61000-4-5
Features
- Burst module (IEC/EN 61000-4-4) up to 5.5kV
- Surge module (IEC/EN 61000-4-5) up to 5kV
- PowerFail module (IEC/EN 61000-4-11)
- Magnetic field tests with optional accessories
- Built-in single phase coupler 300V/16A
- Manual operation from front panel
- USB and GPIB-Bus for remote control
- Monitor outputs for peak voltage and peak current
- Fail inputs
- External variac control 0-10Vdc
- Control of external three-phase couplers
Specifications
TECHNICAL DETAILS | |
Burst Module EFT/N5 as per EN/IEC 61000-4-4 and EN 61000-6-1, -6-2 | |
Test voltage | 200V – 5,500V ± 10%;
100V – 2,750V ± 10% into 50ohm |
Pulse shape | 5/50ns into 50ohm and 1,000ohm |
Rise time tr | 5ns ± 30% into 50ohm;
5ns ± 30% into 1,000ohm |
Pulse width td | 50ns ± 30% into 50ohm;
50ns -15/+100ns into 1,000ohm |
Source impedance | 50ohm |
Polarity | Positive/negative |
COMBINATION WAVE/SURGE | |
Surge Module VCS/N5 as per EN/IEC 61000-4-5 and EN 61000-6-1, -6-2 | |
Voltage (o.c.) | 160V – 5,000V ± 10% |
Pulse front time | 1.2us ± 30% |
Pulse time to half value | 50us ± 20% |
Current (s.c.) | Max. 2,500A ± 10% |
Pulse front time | 8us ± 20% |
Pulse time to half value | 20us ± 20% |
Polarity | Positive/negative/alternating |
Event counter | 1 – 30,000 or endless, selectable |
POWER FAIL, DIPS & INTERRUPTS, VOLTAGE VARIATIONS | |
Power Fail Module PFS/N5 as per EN/IEC 61000-4-11, IEC/EN 61000-4-29 and EN 61000-6-1, -6-2 | |
Channel PF1/PF2 | AC voltage: max. 300V
AC current: max. 16A DC voltage: max. 300V DC current: max. 16A |
Frequency | 16Hz – 500Hz and DC |
Switching time | < 5us into a 100ohm resistive load |
Inrush current | > 500A |
Protection | Both channels are protected against short-circuit conditions. |
GENERAL DATA | |
Interfaces | |
Serial interface | USB |
Parallel interface | IEEE 488, addresses 1 – 30 |
Analog output | 0 – 10VDC to control an external transformer |
CN interface | 15pin SubD connector to control an external coupling network |
Fail inputs | DUT monitoring via Fail1 and Fail2 input (one each) |
Dimensions | |
Dimensions | 19″, 3HU |
Weight | approx. 25kg |
- Basic Standards
- IEC 61000-4-11
- IEC 61000-4-29
- IEC 61000-4-4
- IEC 61000-4-5
- IEC 61000-4-8
- IEC 61000-4-9
- Telecom Standards
- ITU-T K.20
- ITU-T K.21
- ITU-T K.41
- ITU-T K.45
- Product Standards
- EN 300 386 V1.3.2
- EN 50082-1
- EN 50082-2
- EN 50121
- EN 55024
- FCC 97-270 (Part 68)
- IEC 60255-22-5
- IEC 60255-26
- IEC 61326-1
- IEC 61850-3
- EN 300 329
- EN 300 340
- EN 300 342-1
- EN 301 489-1
- EN 301 489-17
- EN 301 489-24
- EN 301 489-7
- Generic Standards
- EN 61000-6-1
- EN 61000-6-2