EM TEST VSS 500N15.3 Voltage Surge Simulator

Description

VSS 500N15.3 Voltage Surge Simulator with 0.5J Constant Pulse Energy up to 15kV

EM TEST VSS 500N15.3 Voltage Surge Simulator generates high voltage transients as required by IEC 60255-27 with a source impedance of 500 ohm for transient overvoltage tests on “Protection Relays”. The output voltage ranges up to 15 kV, divided into two voltage ranges (up to 6.6kV and up to 15kV). The no-load waveshapes corresponds to IEC-61180-1. For this kind of testing the IEC 60255-27 standard requires a fixed energy of 0.5 J at each test level.

The output of the VSS 500N15.3 is divided into two ranges; the lower range covers the test up to 6.6 kV while the higher range supports testing between 6 kV and 15 kV.

The VSS 500N6 offers the Quick Start test routine where parameters can be changed on-line during the test. Fail inputs are offered as standard for DUT monitoring purposes.

 

MODEL OVERVIEW

VSS 500N15.1 – 15kV Voltage Surge Simulator with 0.5J Constant Pulse Energy as per IEC 60255-5

Features

  • Test pulse 1.2/50 µs up to 15 kV as per EN/IEC 60255-27
  • 7 different test levels
  • Source impedance 500 ohm
  • Constant energy of 0.5 J at each test level
  • Peak voltage measurement
  • Peak current measurement
  • Interlock
  • Warning lamp control
  • Manual operation
  • USB (optical link) and GPIB interface
Download Data Sheet
Print
Email

Specifications

TECHNICAL DETAILS
VOLTAGE SURGE SIMULATOR, PULSE 1.2/50US AS PER EN/IEC 60255-27
TECHNICAL DATA
Voltage (o.c.) 250 V – 15,000 V ± 10 %
Front time 1.2 µs ± 30 %
Time to half value 50 µs ± 20 %
Source impedance 500 ohm
Energy 0.5 J ± 0.05 J at each test level
Test level 0.55, 1.0, 3.0, 5.0, 6.6, 9.3 and 14 kV
Pulse forming As per IEC 60255-27, (replaces -5)
Polarity Positive, negative or alternating
Event counter 1 – 30,000 or endless
OUTPUT
Direct Via HV connector; Zi = 500 ohm

To connect external test boxes or devices

Output 1 500V – 6,600 V
Output 2 6,600 – 15,000 V
INTERFACE
Optical interface Opto link, 3 m cable

USB A connector

Parallel interface IEEE 488, addresses 1 to 30
GENERAL DATA
Dimensions, weight 19″/6 HU, 24.9 kg
Supply voltage 115/230 V +10/-15 %
Fuses 2xT2 AT (230 V) or 2xT4 AT (115 V)
Stay informed