Tektronix DPO2024 4 Channel 200 MHz Digital Oscilloscope 1 GSa/s
Description
Offering up to 200 MHz bandwidth and 1 GS/s sample rate, the MSO/DPO2000 Mixed Signal Oscilloscope Series delivers advanced debug features at an entry-level price. With up to 20 channels for analyzing analog and digital signals, you can quickly find and diagnose problems in complex designs. To capture long windows of signal activity while maintaining fine timing resolution, the MSO/DPO2000 offers a deep record length of 1Mpoints standard on all channels.
With Wave Inspector® controls for rapid waveform navigation and automated serial and parallel bus analysis, the MSO/DPO2000 Oscilloscope Series from Tektronix provides the feature-rich tools you need to simplify and speed debug of your complex design.
Features
- 200, 100 MHz Bandwidth Models
- 2 and 4 Analog Channel Models
- 16 Digital Channels (MSO Series)
- 1 GS/s Sample Rate on All Channels
- 1 Megapoint Record Length on All Channels
- 5,000 wfm/s Maximum Waveform Capture Rate
- Suite of Advanced Triggers
Specifications
Characteristic |
MSO2012 DPO2012 |
MSO2014 DPO2014 |
MSO2024 DPO2024 |
Input Channels |
2 |
4 |
4 |
Analog Bandwidth (-3 dB)*1 |
100 MHz |
100 MHz |
200 MHz |
Calculated Rise Time |
3.5 ns |
3.5 ns |
2.1 ns |
Hardware Bandwidth Limits |
20 MHz |
||
Input Coupling |
AC, DC, GND |
||
Input Impedance |
1 MΩ ±2%, 11.5 pF ±2 pF |
||
Input Sensitivity Range |
2 mV/div to 5 V/div |
||
Vertical Resolution |
8 bits |
||
Maximum Input Voltage, 1 MΩ |
300 VRMS with peaks ≤ ±450 V |
||
DC Gain Accuracy (with offset set to 0 V) |
±3% for 10 mV/div to 5 V/div ±4% for 2 mV/div to 5 mV/div |
||
Channel-to-Channel Isolation (Any Two Channels at Equal Vertical Scale) |
≥100:1 at ≤100 MHz |
100:1 at ≤200 MHz |
*1 Bandwidth is 20 MHz at 2 mV/div, all models.
Offset Range
Range |
1 MΩ |
2 mV/div to 200 mV/div |
±1 V |
>200 mV/div to 5 V/div |
±25 V |
Vertical System Digital Channels
Characteristic |
All MSO2000 Models |
Input Channels |
16 Digital (D15 to D0) |
Thresholds |
Threshold per set of 8 channels |
Threshold Selections |
TTL, CMOS, ECL, PECL, User Defined |
User-defined Threshold Range |
±20 V |
Maximum Input Voltage |
±40 V |
Threshold Accuracy |
±(100 mV +3% of threshold setting) |
Maximum Input Dynamic Range |
80 Vpk-pk (threshold setting dependent) |
Minimum Voltage Swing |
500 mVpk-pk |
Input Impedance |
101 kΩ |
Probe Loading |
8 pF |
Vertical Resolution |
1 bit |
Horizontal System Analog Channels
Characteristic |
MSO2012/2014 DPO2012/2014 |
MSO2024 DPO2024 |
Maximum Sample Rate (all channels) |
1 GS/s |
|
Maximum Record Length (all channels) |
1 Mpoints |
|
Maximum Duration of Time Captured at Highest Sample Rate (all channels) |
1 ms |
|
Time-base Range (s/div) |
4 ns to 100 s |
2 ns to 100 s |
Time-base Delay Time Range |
-10 divisions to 5000 s |
|
Channel-to-Channel Deskew Range |
±100 ns |
|
Time-base Accuracy |
±25 ppm |
Horizontal System Digital Channels
Characteristic |
All MSO2000 Models |
Maximum Sample Rate (when using any of channels D7-D0) |
1 GS/s (1 ns resolution) |
Maximum Sample Rate (when using any of channels D15-D8) |
500 MS/s (2 ns resolution) |
Maximum Record Length (all channels) |
1 Mpoints |
Minimum Detectable Pulse Width |
5 ns |
Channel-to-Channel Skew |
2 ns typical |
Trigger System
Characteristic |
Description |
Main Trigger Modes |
Auto, Normal, and Single |
Trigger Coupling |
DC, HF reject (attenuates >85 kHz), LF reject (attenuates <65 kHz), noise reject (reduces sensitivity) |
Trigger Holdoff Range |
20 ns to 8 s |
Trigger Signal Frequency Counter |
Provides a higher accuracy means of identifying the frequency of trigger signals. Trigger Signal Frequency counter resolution is 6 digits. |
Trigger Sensitivity
Characteristic |
Description |
Internal DC Coupled |
0.4 divisions from DC to 50 MHz 0.6 divisions > 50 MHz to 100 MHz 0.8 divisions > 100 MHz to 200 MHz |
External (Auxiliary Input) |
200 mV from DC to 100 MHz, 1X attenuation |
Trigger Level Range
Characteristic |
Description |
Any Channel |
±4.92 divisions from center of screen |
External (Auxiliary Input) |
±6.25 V, 1X attenuation ±12.5 V, 10X attenuation |
Trigger Modes
Mode |
Description |
Edge |
Positive or negative slope on any channel or front-panel auxiliary input. Coupling includes DC, AC, HF reject, LF reject, and noise reject. |
Pulse Width |
Trigger on width of positive or negative pulses that are >, <, =, or ≠ a specified period of time. |
Runt |
Trigger on a pulse that crosses one threshold but fails to cross a second threshold before crossing the first again. |
Logic |
Trigger when any logical pattern of channels goes false or stays true for specified period of time. Any input can be used as a clock to look for the pattern on a clock edge. Pattern (AND, OR, NAND, NOR) specified for all analog and digital input channels defined as High, Low, or Don’t Care. |
Setup and Hold |
Trigger on violations of both setup time and hold time between clock and data present on any of the input channels. |
Rise/Fall Time |
Trigger on pulse edge rates that are faster or slower than specified. Slope may be positive, negative, or either. |
Video |
Trigger on line number, all lines, odd, even, or all fields on NTSC, PAL, and SECAM video signals. |
I2C (optional) |
Trigger on Start, Repeated Start, Stop, Missing ACK, Address (7 or 10 bit), Data, or Address and Data on I2C buses up to 3.4 Mb/s. |
SPI (optional) |
Trigger on SS, MOSI, MISO, or MOSI and MISO on SPI buses up to 10.0 Mb/s. |
CAN (optional) |
Trigger on Start of Frame, Frame Type (data, remote, error, overload), Identifier (standard or extended), Data, Identifier and Data, End of Frame, Missing ACK, or Bit Stuffing Error on CAN signals up to 1 Mb/s. Data can be further specified to trigger on ≤, <, =, >, ≥, or ≠ a specific data value. User-adjustable sample point is set to 50% by default. |
RS-232/422/485/UART (optional) |
Trigger on Tx Start Bit, Rx Start Bit, Tx End of Packet, Rx End of Packet, Tx Data, Rx Data, Tx Parity Error, and Rx Parity Error. |
LIN (optional) |
Trigger on Sync, Identifier, Data, Identifier and Data, Wakeup Frame, Sleep Frame, Errors such as Sync, Parity, or Checksum Errors. |
Parallel (available on MSO models only) |
Trigger on a parallel bus data value. |
Acquisition Modes
Mode |
Description |
Sample |
Acquire sampled values. |
Peak Detect |
Captures glitches as narrow as 3.5 ns at all sweep speeds. |
Averaging |
From 2 to 512 waveforms included in average. |
Roll |
Scrolls waveforms right to left across the screen at sweep speeds slower than or equal to 40 ms/div. |
Waveform Measurements
Measurement |
Description |
Cursors |
Waveform and Screen. |
Automatic Measurements |
29, of which up to four can be displayed on-screen at any one time. Measurements include: Period, Frequency, Delay, Rise Time, Fall Time, Positive Duty Cycle, Negative Duty Cycle, Positive Pulse Width, Negative Pulse Width, Burst Width, Phase, Positive Overshoot, Negative Overshoot, Peak to Peak, Amplitude, High, Low, Max, Min, Mean, Cycle Mean, RMS, Cycle RMS, Positive Pulse Count, Negative Pulse Count, Rising Edge Count, Falling Edge Count, Area and Cycle Area. |
Gating |
Isolate the specific occurrence within an acquisition to take measurements on, using either the screen, or waveform cursors. |
Waveform Math
Characteristic |
Description |
Arithmetic |
Add, subtract, and multiply waveforms. |
FFT |
Spectral magnitude. Set FFT Vertical Scale to Linear RMS or dBV RMS, and FFT Window to Rectangular, Hamming, Hanning, or Blackman-Harris. |
Software
Product |
Description |
NI LabVIEW SignalExpress™ Tektronix Edition |
A fully interactive measurement software environment optimized for the MSO/DPO2000 Series, enables you to instantly acquire, generate, analyze, compare, import, and save measurement data and signals using an intuitive drag-and-drop user interface that does not require any programming. Standard MSO/DPO2000 Series support for acquiring, controlling, viewing, and exporting your live signal data is permanently available through the software. The full version (SIGEXPTE) adds additional signal processing, advanced analysis, mixed signal, sweeping, limit testing, and user-defined step capabilities and is available for a 30-day trial period standard with each instrument. |
OpenChoice®Desktop |
Enables fast and easy communication between a Windows PC and the MSO/DPO2000 Series. Transfer and save settings, waveforms, measurements, and screen images. Included Word and Excel toolbars automate the transfer of acquisition data and screen images from the oscilloscope into Word and Excel for quick reporting or further analysis. |
IVI Driver |
Provides a standard instrument programming interface for common applications such as LabVIEW, LabWindows/CVI, Microsoft .NET, and MATLAB. |
eScope |
Enables control of the MSO/DPO2000 Series over a network connection through a standard web browser. Simply enter the IP address or network name of the oscilloscope and a web page will be served to the browser. |
Display Characteristics
Characteristic |
Description |
Display Type |
7 in. (180 mm) liquid crystal TFT color display. |
Display Resolution |
480 horizontal × 234 vertical pixels (WQVGA). |
Waveform Styles |
Vectors, Dots (In Video Trigger mode), Variable Persistence, Infinite Persistence. |
Graticules |
Full, Grid, Cross Hair, Frame. |
Format |
YT and XY. |
Maximum Waveform Capture Rate |
Up to 5,000 wfm/s. |
Input/Output Ports
Port |
Description |
USB 2.0 High-speed Host Port |
Supports USB mass storage devices and keyboards. One port available on front panel. |
USB 2.0 High-speed Device Port |
Rear-panel connector allows for communication/control of oscilloscope through USBTMC or GPIB with a TEK-USB-488, and direct printing to all PictBridge-compatible printers. |
LAN Port |
RJ-45 connector, supports 10/100Base-T (requires DPO2CONN). |
Video Out Port |
DB-15 female connector, connect to show the oscilloscope display on an external monitor or projector (requires DPO2CONN). |
Auxiliary Input |
Front-panel BNC connector. Input Impedance 1 MΩ ±2%. Max input 300 VRMS CAT II with peaks ≤ ±450 V. |
Probe Compensator Output |
Front-panel pins Amplitude: 5 V Frequency: 1 kHz |
Kensington Style Lock |
Rear-panel security slot connects to standard Kensington-style lock. |
Power Source
Characteristic |
Description |
Power Source Voltage |
100 to 240 V ±10% |
Power Source Frequency |
45 to 65 Hz (90 to 264 V) 360 to 440 Hz (100 to 132 V) |
Power Consumption |
80 W maximum |
Optional TekVPI® External Power Supply (119-7465-xx) |
Output Voltage: 12 V Output Current: 5 A Power Consumption: 50 W |
Physical Characteristics
Dimensions |
mm |
in. |
Height |
180 |
7.1 |
Width |
377 |
14.9 |
Depth |
134 |
5.3 |
Weight |
kg |
lb. |
Net |
3.6 |
7.9 |
Shipping |
6.2 |
13.7 |
Rackmount Configuration |
4U |
|
Cooling Clearance |
2 in. (50 mm) required on left side and rear of instrument |
Environmental
Characteristic |
Description |
Temperature |
|
Operating |
0 ºC to +50 ºC |
Non operating |
-40 ºC to +71 ºC |
Humidity |
|
Operating |
High: 30 ºC to 50 ºC, 5% to 60% Relative Humidity Low: 0 ºC to 30 ºC, 5% to 95% Relative Humidity |
Nonoperating |
High: 30 ºC to 55 ºC, 5% to 60% Relative Humidity Low: 0 ºC to 30 ºC, 5% to 95% Relative Humidity |
Altitude |
|
Operating |
3,000 meters (9,843 feet) |
Nonoperating |
12,000 meters (39,370 feet) |
Random Vibration |
|
Operating |
0.31 GRMS from 5 to 500 Hz, 10 minutes each axis, 3 axes, 30 minutes total |
Non operating |
2.46 GRMS from 5 to 500 Hz, 10 minutes each axis, 3 axes, 30 minutes total |
Regulatory |
|
Electromagnetic Compatibility |
EC Council Directive 2004/108/EC |
Safety |
UL61010-1:2004; CAN/CSA C22.2 No. 61010.1-04; EN61010-1:2001; Complies with the Low Voltage Directive 2004/108/EC for Product Safety. |