Teseq CDN 3043 3-Phase Coupling/Decoupling Network

Description

Teseq CDN 3043 Three Phase Coupling/Decoupling Network designed for use as extension to NSG 3040 multifunction generator series.

Dedicated EUT terminals allow direct and easy EUT connection; no specific adapter plugs are required. A safety isolated terminal key is part of the delivery.

EUT power ON and OFF can be controlled manually from the NSG 3040 front panel, as well as automatically from WIN 3000 software. EUT power will also switch OFF automatically in case of overload.

High value shielded High Voltage safety plugs are used to interconnect the high voltage pulse output from NSG 3040 multi-function generator to CDN input.

Features

  • Compliant with IEC 61000-4-4 & 5
  • Universal terminals – no special plugs required
  • Safety HV plugs and terminals
  • Manual/programmable EUT power ON/OFF control
  • Short term overstress capability
Download Data Sheet
Print
Email

Specifications

Parameter Value
CDN 3043-Bxx
CDN 3043-Sxx
CDN 3043-Cxx
For EFT/Burst
For combined wave surge (CWS)
Combined EFT and CWS
EUT voltage 30 to 480 VAC RMS, 0 to 65 Hz
0 to 125 VDC – full current range
0 to 225 VDC – up to 7 A
EUT current 32 A continous CDN 3043-x32 series
45 A for ca. 30 mn
Burst voltage Up to 4800 V
Surge voltage Up to 4400 V
Terminals Screw terminals, rated for 32 A
Size 449 x 310 x 565 mm
Weight 40 kg approx.