Teseq NSG 434 ESD Simulator 8.8 kV

Description

Teseq NSG 434 ESD Simulator 8.8 kV is designed to sit comfortably in the operator’s hand, with current operating conditions constantly displayed and clearly visible.

NSG 434’s microprocessor-based controller and multifunction keypad provide the operator with instant access to its comprehensive range of built-in functions. The LCD panel continuously displays the operating status and all user-selected test parameters.

NSG 434 has its own internal, battery operated high voltage generator. In addition to the pre-programmed, standard IEC pulses, the user can create custom tests using single or repetitive discharges with selectable rates and manual or automatic polarity switching.

Features

  • Programmable discharge voltage
  • 200 V to 8.8 kV
  • Stabilized charging voltage
  • Pre-programmed IEC/EN 61000-4-2
  • test settings up to level 3
  • True air-discharge breakdown
  • voltage
  • Battery powered
  • Compact and lightweight design
  • Comfortable use and convenient
  • operation
Download Data Sheet
Print
Email

Specifications

Parameter Value
Basic set Carrying case with: electrostatic discharge simulator, battery pack, battery charger 100 to 240 VAC, discharge network 150 pF/330 Ω, air- and contact-discharge tips, grounding cable, user manual
Pulse
network
Network 150 pF/330 Ω as per IEC/EN 61000-4-2 (included)
Range of R/C networks for other standards: R = 0 Ω to 10 kΩ; C = 60 to 500 pF
Discharge voltage Air-discharge: 200 V to 8.8 kV (in 100 V steps)
Contact-discharge: 200 V to 6.6 kV (in 100 V steps)
Charging voltage measurement kV, accuracy better than ±5% (stabilized); measurement and display of true air flashover voltage
Discharge detection Indicated by an oval around the kV symbol, also acoustically in the ‘Single’ operating mode (air-discharge only)
Holding time >5 s (charging voltage … ±5%)
Operating modes Single/repetitive =
pulse counter: 0 to 9999, preselect counter 0 to 9999
Repetition: 0.5/1/5/10/20 or 25 Hz (air); 0.5/1/5 or 10 Hz (contact); continuous operation
Displaytext LCD panel showing: charging voltage, discharge voltage, polarity, air-/contact-discharge, counter/preselect counter content, soft-key functions, battery monitor
Weight NSG 434 with battery: 1.2 kg (2.6 lbs) approx.
Ambient conditions Operation: +5 to +40°C, 20 to 80 % r.h. (non-condensing), 68 to 106 kPa