EM TEST UCS 500N5P Power Fail Simulator

Description

UCS 500N5P Power Fail Simulator for Dips and Short Interruptions as per IEC/EN 61000-4-11 and 61000-4-29

EM TEST UCS 500N5P Power Fail Simulator is a fully compliant simulator as per IEC 61000-4-11 and IEC 61000-4-29 with the best benefit-to-cost ratio yet upgradeable to include other test abilities such as EFT/Burst and Combination Wave Surge.

The UCS 500N5P includes everything necessary to conduct fully compliant tests. The power mains supply for the controls and for the DUT is separate to render it more flexible to use with different DUT supply voltages.

Pre-programmed Standard Test routines allow users to change parameters on-line during their test to evaluate the susceptibility level of an individual DUT.

 

MODEL OVERVIEW

UCS 500N5 Transient Generator for EFT/Burst, Surge & Power Fail up to 5.5 kV as per IEC/EN 61000-4-4, 61000-4-5 and 61000-4-11
UCS 500N5E Electrical Fast Transient /Burst Generator up to 5.5 kV as per IEC/EN 61000-4-4
UCS 500N5V Combination Wave (Surge only) Generator up to 5kV as per IEC/EN 61000-4-5

Features

  • Fully compliant to IEC 61000-4-11 Ed.2 as well as IEC 61000-4-29
  • Voltage/current up to 300V/32A AC/DC
  • Switching time between 1us and 5us
  • Inrush current > 500A
  • Monitor outputs for voltage and current
  • Fail inputs
  • External variac control 0-10Vdc
  • Magnetic field tests with optional accessories
  • Manual operation from front panel
  • USB and GPIB-Bus for remote control
Download Data Sheet
Print
Email

Specifications

TECHNICAL DETAILS
As per EN/IEC 61000-4-11, IEC/EN 61000-4-29 and EN 61000-6-1, -6-2
Channel PF1/PF2 AC voltage: max. 300V

AC current: max. 32A

DC voltage: max. 300V

DC current: max. 32A

Frequency AC: 16Hz – 500Hz and DC
Switching time 1us to 5us into a 100ohm resistive load
Inrush current > 500A
Protection Both channels are protected against short-circuit conditions.
OUTPUTS
DUT terminals L,N and PE
CRO trigger 5V trigger signal oscilloscope
INTERFACES
Serial interface USB
Parallel interface IEEE 488, addresses 1 – 30
Analog output 0 – 10VDC to control an external transformer
Fail inputs DUT monitoring via Fail1 and Fail2 input (one each)
GENERAL DATA
Dimensions 19″, 3HU
Weight approx. 15kg
MAINS
Supply voltage 115V/230VAC +10%/-15%
Power approx. 75W
Frequency 50/60Hz
Fuses 2xT2AT (230V) or 2xT4AT (115V)