EM TEST VSS 500N12-series Voltage Surge Simulator
Description
VSS 500N12-series – 12kV Voltage Surge Simulator for voltage withstand testing as per IEC 61180-1, IEC 61180-2 and IEC 60664
EM TEST VSS 500N12-series Voltage Surge Simulator is specifically designed to test insulation material, components, sockets, connectors, cables and many more items to their voltage withstand capability. Testing the voltage withstand capability by means of a transient test pulse is the most common alternative to tests using A.C. or D.C. voltages.
The VSS 500N12-series has built-in voltage and current monitors and the Spark Over Detection which offers detailed test results to judge the quality of the DUT. Safety precautions are taken to assure safe operation at this high voltage level.
The VSS 500N12-series provides interlock and warning lamp control. By means of an optional test box the operator can be further protected to avoid direct contact with high voltage and to avoid harm from exploding components or fragments of them when failing.
MODEL OVERVIEW
VSS 500N12 12 kV Pulse generator, 1.2/50µs, as per IEC 60060, 500 ohm impedance and low energy
VSS 500N12.1 12 kV Pulse generator, 1.2/50µs, as per IEC 60335, 40 ohm impedance and low energy
VSS 500N12.6 12 kV Pulse generator, 1.2/50µs, as per IEC 61180-1, 50 ohm impedance
VSS 500N12.7 12 kV Pulse generator, 1.2/50µs, as per IEC 60950, 13 ohm + 25 ohm impedance, special pulse circuit, see datasheet VSS 500N12.7
Features
- Surge voltage up to 12kV
- Wave form 1.2/50µs
- Built-in source impedance
- Interlock
- Warning lamp control
- Voltage/current measurement
- Spark-over detection
- Manual operation
- USB (optical link) and GPIB interface
Specifications
TECHNICAL SPECIFICATIONS | |
VOLTAGE SURGE SIMULATOR, PULSE 1.2/50µs | |
Voltage (o.c.) | 500V – 12,000V ± 10% |
Pulse front time | 1.2µs ± 30% |
Pulse time to half | value 50µs ± 20% |
Current (s.c.) | Max. 24A ± 10% |
Int. resistor | 500 ohm ± 10 % (VSS 500N12)
40 ohm ± 10 % (VSS 500N12.1) 50 ohm ± 10 % (VSS 500N12.6) |
Polarity | Positive/negative/alternating |
Event counter | 1 – 30,000 or endless |
TRIGGER | |
Automatic | Automatic pulse release |
Manual | Single pulse release |
External | External pulse release |
CRO trigger | 5V trigger signal for oscilloscope |
Synchronization | 0° – 360°, resolution 1° |
Repetition rate | Max. 1Hz (1s – 999s) |
OUTPUT | |
Direct | Via HV connector; Zi = 500ohm or as per selected model(40 ohm or 50 ohm) |
INTERFACE | |
Optical interface | Opto link, 3m cable
USB A connector |
Parallel interface | IEEE 488, addresses 1 – 30 |
GENERAL DATA | |
Dimensions | 19″/3HU |
Weight | approx. 15kg |
Supply voltage | 115/230V +10/-15% |
Fuses | 2 x 2 AT (230V) or 2 x 4 AT (115V) |