AE Techron 7548RLY Power Amplifier/Current Source

Description

AE Techron 7548RLY Power Amplifier/Current Source is designed to meet the demanding requirements of the power utility industry. Capable of outputting up to 100 amperes peak current, the 7548RLY is powerful enough to put protection relays, fuses and other critical components through a full range of tests. The low noise floor, low distortion and minimal phase error of the 7548RLY make it the ideal amplifier for power grid modeling.

Features

  • Up to 100 Ap (70A RMS) output power at 0.5 ohms
  • Up to 195 Vp output voltage
  • Standard transconductance of 20 ±0.2% from short to 1-ohm load
  • Residual noise of less than 2.5 mA peak, from 40 Hz to 600 Hz
Download Data Sheet
Print
Email

Specifications

Performance
Frequency Response DC to 10 kHz driving a 0.25-ohm load
Maximum Output Current 70 Arms, 100 Apeak, 0.5 ohms
Maximum Output Voltage 195 Vpeak
Maximum Output Power Dependent on load and frequency
Load Constraint for Maximum Output 0.5 ohms + 200 microhenries
Output Offset Current Less than 10.0 milliamperes DC peak
Unit to Unit Phase Error ± 0.1 degrees at 60 Hz
Residual Noise Less than 2.5 milliamperes peak (40Hz – 600Hz)
Out Accuracy Less than ±1%
Input Characteristics
Balanced with ground Three terminal barrier block connector 20 k ohm differential
Unbalanced BNC connector, 10 k ohm single ended
Standard Gain Switchable 10 amps/volt (+0.2%) or 2 amps/volt (+0.2%)
Standard Transconductance 20 ±0.2% from short to 1-ohm loads
Max Input Voltage ± 10 V balanced or unbalanced
Input Impedance 20 kOhm differential
Input Sensitivity 3.0V input for 3800W output into 1 ohm (adjustable)
Common Mode Rejection Ratio -58 dB minimum, 40-600Hz
Display, Control, Status, I/O
Front Panel LED Displays Indicate Run, Ready, Standby, Stop, and Fault conditions in the output stage
LCD Display Lists type of fault condition and gives suggested corrective action
Soft-Touch Switches for Run (Enable), Stop, Reset
User Configurable LCD display can be configured for up to four simultaneous displays reporting one, two or all four of the following: Voltage Peak, Voltage RMS, Current Peak, and Current RMS
Back Panel Power Connection NEMA-style locking receptacle; matching AC connector also included
Signal Output 4-position terminal barrier block
Signal Input User-selectable Unbalanced BNC or Balanced Barrier Strip
Interlock Connector 25-pin D-sub connector used for amplifier control and status applications; also used in multi-amplifier applications
Communication Capabilities
Current Monitor ± 1V / 20A ±1%
Reporting System Fault, OverTemp, Over Voltage, Overload
Control Force to Standby; Reset after a fault
Protection
Over/Under Voltage ± 10% from specified supply voltage amplifier is forced to Standby
Over Current Breaker protection on both main power and low voltage supplies
Over Temperature Separate Output transistor, heat sink, and transformer temperature monitoring and protection
Physical Characteristics
Chassis Black powder-coat chassis with all aluminum construction; designed for stand-alone or rack-mounted operation. The amplifier occupies five EIA 19-inch-wide rack units
Weight 110 lbs. (50 kg)
AC Power Three-phase, 208 VAC ±10%, 47-60 Hz, 20A AC service (400 VAC ±10%, 15A version available). A toggle switch circuit breaker opens all legs of the AC mains on excess current demand.
Operating Temperature 10°C to 50°C (50°F to 122°F), Maximum Output Power de-rated above 30°C (86°F)
Humidity 70% or less, non-condensing
Cooling Forced air-cooling from front to back through removable filters via four 100 ft3/min. fans. No space is required between rack-mounted amplifiers. Air filters are removable from the rear via one fastener per side and may be eliminated if cabinet filtration is provided.
Dimensions 19 in. x 22.8 in. x 8.75 in. (48.3 cm x 57.9 cm x 22.3 cm).