EM TEST AIF 503N63.1 3-Phase Flicker Impedance 63A
Description
EM TEST AIF 503N63.1 3-Phase Flicker Impedance 63A offers two different impedance values. Zref is used as the flicker impedance as per IEC/EN 61000-3-3 and specified by IEC 60725. Ztest is used for equipment being intended for connection to a public low-voltage distribution system having a service current supply capability of 100A and above as specified in IEC/EN 61000-3-11.
MODEL SERIES OVERVIEW
AIF 503N-series Three-phase Flicker Impedances up to 75A
Features
- Impedances values as per IEC/EN 61000-3-3 & IEC/EN 61000-3-11
- Non-saturable air coil inductors
- Line voltage 3 x 400V
- Line current up to 75A per phase (depending on model)
- Automatic overcurrent protection
- Status indication by LED
- Remote-controlled by DPA 503N
Specifications
Line current | 63A per phase |
3-ph output | High-current safety connectors 80A |
1-ph output | One for each phase (L1,L2,L3),
connector country specific |
1-ph low current | Connector on phase L1,
Range: 100mA – 10A |
Mains supply | 115/230V, 200W, 50/60Hz |
Housing | 19″, 38HU, moveable rack on castors |
Weight | 230kg |
Size | 1800mm x 600mm x 800mm |