EM TEST AIF 503N-series 3-Phase Flicker Impedances 16A – 75A

Description

EM TEST AIF 503N-series 3-Phase Flicker Impedances 16A – 75A unifies the required impedance values for two applications (except for the AIF 503N16 model). It offers a flicker impedance with the values in accordance with IEC 60725 for flicker measurements as per IEC/EN 61000-3-3. It also includes a second impedance used for flicker measurements on DUTs with a rated current above 16A which are intended to be connected to a public low-voltage distribution system with a service current of more than 100A.

The inductors used are non-saturable air coils and are manually matched to the specified value.

The AIF 503N-series is used along with the DPA503N for single phase as well as for three phase measurements.

 

MODEL OVERVIEW

AIF 503N16 Three-Phase Flicker Impedance 16A
AIF 503N32.1 Three-Phase Flicker Impedance 32A
AIF 503N63.1 Three-Phase Flicker Impedance 63A
AIF 503N75.1 Three-Phase Flicker Impedance 75A

Features

  • Impedances values as per IEC/EN 61000-3-3 & IEC/EN 61000-3-11
  • Non-saturable air coil inductors
  • Line voltage 3 x 400V
  • Line current up to 75A per phase (depending on model)
  • Automatic overcurrent protection
  • Status indication by LED
  • Remote-controlled by DPA 503N
Download Data Sheet
Print
Email

Specifications

COMMON DATA (ALL MODELS)
GENERAL
Configuration 3-phase (3L/N/PE)
Line voltage 3 x 400V ± 10%
Line frequency 47Hz – 63Hz
Bypass for harmonics Remote controlled switch
Interface 15-pin SubD (DPA 503N control)
Indicators LED’s indicating the active status (Bypass or impedance in use).
IMPEDANCE
Zref at 50Hz RA = 0.24ohm, XA = 0.15ohm

RN = 0.16ohm, XN = 0.10ohm

Ztest at 50Hz RA = 0.15ohm, XA = 0.15ohm

RN = 0.10ohm, XN = 0.10ohm (not available in AIF 503N16)

Accuracy Better than 3%

Each inductor is a non-saturable air coil, matched manually to the specified value.

AMBIENT CONDITIONS
Temperature 0°C to 35°C
Relative humidity 80%, non-condensing
Atmospheric pressure 86 kPa (860 mbar) to 106 kPa (1,060 mbar)
Cooling fans, active cooling control