EM TEST VDS 200Q-series Voltage Drop Simulator
Description
EM TEST VDS 200Q-series 4-Quadrant Voltage Drop Simulator is used to simulate various battery supply waveforms recommended by ISO and Automotive OEM standards.
The VDS 200Q-series simulators serve as a powerful DC voltage supply for the DUT during the tests with automotive transients. The VDS 200Q-series handles all three supply voltage categories (12 V, 24 V, and 42 V). The current capability ranges up to 100A (depending on the model and your application).
The VDS 200Q has an integrated DC amplifier with a frequency range up to 150 (180) kHz. It allows us to generate dips and drops, short interruptions along with a variety of voltage variations representing various phenomena being measured on a wiring harness.
The VDS 200Q-series are four quadrant bipolar amplifiers capable of absorbing energy fed back from the DUT up to nominal current.
MODEL OVERVIEW
VDS 200Q10 – Voltage Drop Simulator, 60 V / 10 A
VDS 200Q25 – Voltage Drop Simulator, 60 V / 25 A
VDS 200Q25.1 – Voltage Drop Simulator, 80 V (60 V) / 20 A (25 A)
VDS 200Q50 – Voltage Drop Simulator, 60 V / 50 A
VDS 200Q50.1 – Voltage Drop Simulator, 80 V (60 V) / 40 A (50 A)
VDS 200Q100 – Voltage Drop Simulator, 60 V / 100 A
VDS 200Q100.1 – Voltage Drop Simulator, 80 V (60 V) / 80 A (100 A)
Features
- Voltage up to 60 V, 2 ranges
- Expandable to 77 V
- Current up to 200 A (600 A peak)
- Four quadrant operations
- Fast step response time
- Low output impedance
- High Bandwidth
- Temperature-controlled air cooling
- Pre-programmed test routines to simulate various supply waveforms
- Input for external signal generator to produce most complex waveforms
- Software for Windows Vista & Windows 7
- USB and GPIB interfaces
Specifications
GENERAL | |
Source impedance | Zi = <10 mohm |
Operation | 4 – quadrant, bipolar operation |
Current limiter | (not available for VDS 200Q10) 3x Imax: allows an inrush current of three times nominal current for 200 ms before the current limiter starts 3x ictrl: allows an inrush current of three times the programmed current for 200 ms before the current limiter starts Peak OFF: no inrush current above the set current value |
Compensation | STD: DC – 40 kHz
HF: DC – >=150 kHz CAP: DC – 3kHz |
Recovery | >90% of excursion within 25 us |
Output rise time | typ. <10 us, <3 us (high freq.) |
Ripple voltage | Ur <10 mVp-p, frequency min. 400 Hz |
Control | Analog In |
Cooling | temperature-controlled air cooling |
Protection | Thermal-Magnetic Circuit Breakers Depending on VDS 200Q model |
TRIGGER | |
Automatic | Automatic release of the events |
Manual | Manual release of a single pulse |
External | External release of a single pulse |
OUTPUT | |
DUT Supply +/- | Safety laboratory or high current connectors |
Ext. trigger | 5-15 V TTL, BNC connector |
CRO Trigger | 5 V TTL-signal for oscilloscope |
INTERFACE | |
Interfaces | USB
Ethernet (for optional AutoWave) IEEE 488, addresses 1 – 30 |
Remote control | To connect an external signal generator (10 kohm): -10 V – +10 V / 0 – 150 kHz (180 kHz for VDS 200Q10) |