EM TEST VCS 500N8.5 Combination Wave Simulator
Description
EM TEST VCS 500N8.5 Combination Wave Simulator generates surge voltages up to 8kV and surge currents up to 4kVA as required for by IEC/EN 61000-4-5 (e.g. overvoltage protection for telecommunication devices).
The VCS 500N8.5 offers an integrated coupling/decoupling network, either single phase or three-phase, with a nominal current of 16A per phase. In case higher current is required its capability can be extended by an external CDN to achieve nominal currents of 100A and more per phase.
MODEL SERIES OVERVIEW
VCS 500N8-series Combination Wave Simulator with integrated 1-Phase or 3-Phase Coupling/Decoupling Network
Features
- Surge voltage up to 8kV
- Surge current up to 4kA
- Voltage/current monitors
- Built-in CDN 1-phase or 3-phase up to 3 x 690V and 32A
- Interlock
- Warning lamp control
- Manual operation
- USB (optical link) and GPIB interface
Download Data Sheet
Print
Email
Specifications
TECHNICAL DETAILS | |
COMBINATION WAVE 1.2/50US – 8/20US | |
Voltage (o.c.) | 250V – 8,000V ± 10% |
Front time | 1.2μs ± 30% |
Time to half value | 50μs ± 20% |
Current (s.c.) | Max. 4,000A ± 10% |
Front time | 8μs ± 20% |
Time to half value | 20μs ± 20% |
Polarity | Positive/negative/alternating |
Event counter | 1 – 30,000 or endless |
TRIGGER | |
Trigger of events | Automatic, manual, external |
CRO trigger | 5V trigger signal for oscilloscope |
Synchronization | 0° – 360°, resolution 1° |
Repetition rate | 3s – 999s, depending on the voltage |
OUTPUT | |
VCS 500N8.5 | Built-in 3-phase coupler 3 x 690VAC / 32A; 50/60Hz
DC: 400V/20A |
INTERFACE | |
Serial interface | USB |
Parallel serial interface | IEEE 488, addresses 1 – 30 |
SAFETY | |
Safety circuit | Control input (24Vdc) |
Warning lamp | Floating output contact |
GENERAL DATA | |
Dimensions – Weight | 19″/9HU, approx. 52kg |
Supply voltage | 115/230V +10/-15% |
Fuses | 2x2AT (230V) or 2x4AT (115V) |