EM TEST Compact NX7 Transient Generator
Description
Compact NX7 – Compact Tester for EFT/Burst, Surge, Power Fail & Ringwave up to 7 kV
EM TEST Compact NX7 Transient Generator is the most versatile tester to address transient and power fail requirements for both international and commercial standards.
The NX7 provides an economical solution for pre-compliance immunity testing as well as full-compliance testing and CE Marking. Its internal single-phase Coupling/Decoupling Network (CDN) can be extended for testing three-phase EUTs by means of an automatically controlled external CDN up to 200 A per phase.
Featuring an easy-to-use color touch screen, the NX7 is a standalone generator which includes everything necessary to perform fully compliant tests.
MODEL OVERVIEW
Compact NX5 Multifunctional Test Generator for Transients (EFT/Burst, Surge & Power Fail) up to 5.5 kV
Compact NX5 Multifunctional Test Generator for Conducted Transients (EFT/Burst, Surge, Telecom-Surge & Power Fail) up to 5.5 kV
Features
- Smallest compact generator with 7″ touch screen
- Burst 5.5 kV, Surge & Ringwave 7.0 kV and Power Fail
- Built-in single phase CDN 400 V/32 A
- Manual front panel operation with setup pictures
- Separate key for START/STOP operation with LED
- Monitor outputs for peak voltage and current
- Opto-Link and Ethernet for remote control
- Control of external three-phase couplers
- Module PowerFail as per IEC/EN 61000-4-8/-11/-29(DC)
- Fail inputs
Specifications
ELECTRICAL FAST TRANSIENTS | |
BURST MODULE | As per IEC/EN 61000-4-4 and EN 61000-6-1, -6-2 |
COMBINATION WAVE / SURGE | |
SURGE MODULE | As per IEC/EN 61000-4-5 and IEC/EN 61000-6-1, -6-2 |
POWER FAIL, DIPS & INTERRUPTIONS, VOLTAGE VARIATIONS | |
POWER FAIL MODULE | As per IEC/EN 61000-4-11, IEC/EN 61000-4-29 and IEC/EN 61000-6-1, -6-2 |
RINGWAVE | |
RINGWAVE MODULE | As per ANSI/IEEE C62.41 and IEC/EN 61000-4-12 |
GENERAL DATA | |
INTERFACES | |
Serial interface | 2 x USB A for memory stick
1 x USB B for service only Opto – Link to USB for remote |
Lan | Ethernet for remote |
Analog output | 0 – 10 VDC to control an external transformer |
Sys.link | 26 pin high density connector to control an external coupling network |
Fail inputs | EUT monitoring via input (one each)
EUT Monitor 1 EUT Monitor 2 |
Ext. Trigger | BNC Ext. Trigger IN
pos slope 5 V |
Ext. Sync Input | Differential input
50 V – 690 VAC 2 x 4 mm MC Safety connectors |
DIMENSIONS AND WEIGHT | |
16 A models | 19″/3 HU, 500 mm deep,
19”/9 HU with telecom module, approx. 30 kg |
32 A models | 19″/6 HU, 500 mm deep,
19” 9 HU with telecom module, approx. 40 kg |
MAINS | |
Supply voltage | 85 V – 264 V |
Frequency | 50/60 Hz |
Power | approx. 75 W |
Fuses | 115 V: 2 x 4 A slow blow
230 V: 2 x 2 A slow blow |
- ANSI/IEEE Standards
- ANSI/IEEE C37.90
- ANSI/IEEE C62.41
- Basic Standards
- IEC 61000-4-11
- IEC 61000-4-12
- IEC 61000-4-29
- IEC 61000-4-4
- IEC 61000-4-5
- IEC 61000-4-8
- IEC 61000-4-9
- Telecom Standards
- ITU-T K.20
- ITU-T K.21
- ITU-T K.41
- ITU-T K.45
- Product Standards
- EN 300 386 V1.3.2
- EN 50121
- EN 55024
- FCC 97-270 (Part 68)
- IEC 60255-22-5
- IEC 61326-1
- IEC 61850-3
- EN 300 329
- EN 300 340
- EN 300 342-1
- EN 301 489-1
- EN 301 489-17
- EN 301 489-24
- EN 301 489-7
- Generic Standards
- EN 61000-6-1
- EN 61000-6-2
- Automotive OEM Specifications
- UNECE R10 (Automotive)