Teseq CIP 9138 Current Injection Probe 4 kHz - 200 MHz
Description
Teseq CIP 9138 Current Injection Probe 4 kHz – 200 MHz has been designed for MIL-STD-461G CS114 with the extended range from 4 kHz to 1 MHz for EUTs intended to be installed on ships or submarines.
The CIP 9138 core material is highly efficient and thermally rugged, thus allowing very high injection levels to be achieved with low RF input power. It can withstand far higher powers than conventional ferrite (up to 1000 W), without changes in the characteristics in relation to the temperature.
The probe material meets the requirements of MIL-STD-461 CS114, ISO 11452-4 and other standards.
The probe performance can be measured using calibration jig PCJ 9202.
Features
- Frequency range from 4 kHz to 200 (400) MHz
- High power handling (up to 1 kW)
- Ideal for automotive BCI testing e.g. MIL-STD-461 CS114, ISO 11452-4 and other standards
Specifications
Frequency range | 4 kHz – 200 (400) MHz |
Window diameter | 43 mm |
Outside diameter | 113 mm |
Width | 110 mm |
Weight | approx. 3.7 kg |
Input connector | Type N |
Max. input power | 1000 W |
Max. time for continuous operation | related to the core temperature |
Rating at 10 kHz / 1000 W | approx. 10 min* |
Rating at 100 kHz / 500 W | approx. 7 min* |
Rating at 150 kHz / 500 W | approx. 5 min* |
Rating at 1 MHz to 400 MHz / 400 W | approx. 3 min* |
Max core temperature | 90 ˚C |
Turns ratio | 1:1 |