Teseq CIP 9138 Current Injection Probe 4 kHz - 200 MHz

Description

Teseq CIP 9138 Current Injection Probe 4 kHz – 200 MHz has been designed for MIL-STD-461G CS114 with the extended range from 4 kHz to 1 MHz for EUTs intended to be installed on ships or submarines.

The CIP 9138 core material is highly efficient and thermally rugged, thus allowing very high injection levels to be achieved with low RF input power. It can withstand far higher powers than conventional ferrite (up to 1000 W), without changes in the characteristics in relation to the temperature.

The probe material meets the requirements of MIL-STD-461 CS114, ISO 11452-4 and other standards.

The probe performance can be measured using calibration jig PCJ 9202.

Features

  • Frequency range from 4 kHz to 200 (400) MHz
  • High power handling (up to 1 kW)
  • Ideal for automotive BCI testing e.g. MIL-STD-461 CS114, ISO 11452-4 and other standards
Download Data Sheet
Print
Email

Specifications

Frequency range 4 kHz – 200 (400) MHz
Window diameter 43 mm
Outside diameter 113 mm
Width 110 mm
Weight approx. 3.7 kg
Input connector Type N
Max. input power 1000 W
Max. time for continuous operation related to the core temperature
Rating at 10 kHz / 1000 W approx. 10 min*
Rating at 100 kHz / 500 W approx. 7 min*
Rating at 150 kHz / 500 W approx. 5 min*
Rating at 1 MHz to 400 MHz / 400 W approx. 3 min*
Max core temperature 90 ˚C
Turns ratio 1:1
Stay informed

To keep you in the loop of our latest news, register now for our email newsletter.