ETS-Lindgren HI-6053 Electric Field Probe 10 MHz – 40 GHz
Description
The battery-powered HI-6053 Electric Field Probe provides broadband frequency coverage and wide dynamic range that satisfies the demands of most test requirements. To take advantage of this capability, the HI-6053 was designed to be single range reading so data can be read continuously over the entire dynamic range. Data values for each axis (X, Y, and Z) can be read individually or summed.
The HI-6053 Electric Field Probe is a fully intelligent sensor enabling fast and accurate EMF measurements with industry-leading performance specifications. Optical coupling to a variety of readout options makes this probe ideally suited for a wide range of field monitoring applications.
The HI-6100 Field Monitor provides manual functions and programmed control via IEEE-488 and RS-232 Serial Data Interfaces. Readings from up to four probes can be displayed simultaneously.
For direct connection to a PC either the HI-4413P Serial or the HI-4413USB interfaces are available.
This field probe incorporates user replaceable/rechargeable batteries.
Features
- Frequency Range: 10 MHz – 40 GHz
- Provides Individual and Summed Axis Values
- Dynamic Range: 2.0 – 800 V/m (single range)
- User replaceable/rechargeable batteries
- A2LA Accredited Calibration Report
- Suitable for MIL Standard Specs:
— MIL-STD 461F Radiated Susceptibility (RS) - Suitable for Automotive Specs:
— SAE J1113/27
— GM9120P
— GM93100GS
— GM9114P - Operates with most 3rd Party Immunity software
Specifications
Electrical Specifications | |
Detection | Isotropic (X, Y, and Z Axis Readings) |
Frequency Response with Correction | 10 MHz to 18 GHz ± 0.9 dB 18 to 40 GHz ± 1.1 dB |
Frequency Range | 10 MHz to 40 GHz |
Frequency | 10 to 100 MHz + 3.0, -4.0 dB 100 MHz to 1 GHz +3.0, -0.50 dB 1 to 18 GHz +4.0, -2.0 dB 18 to 40 GHz +3.5, -4.5 dB |
Dynamic Range | 2 to 800 V/m (Single Range) |
Resolution | 0.01 V/m |
Isotropic Deviation | ±1.0 dB < 18 GHz |
Linearity | ±0.5 dB @ 1 GHz |
Sample Rate (Typical) | HI-4413P Serial Modem – 44 samples/second maximum HI-6100 Field Monitor – 70 samples/second maximum |
Overload Withstand | >1500 V/m CW |
Physical Specifications | |
Physical Interface | FSMA Connectors Duplex Optical Fiber (200 micron multimode) |
Battery | Rechargeable NiMH AAA |
Operating Temperature Range | 10° to 40° C, 5-95% Relative Humidity Non-condensing |
Battery Life | >30 Hours Continuous |
Battery Charger | 100-240 VAC Universal Input 3-Hour Charge from Full Depletion |
Weight | 0.36 kg (12.64 oz) |
Mounting | 1/4-20 UNC Internal Thread |
Dim. L. | 43.8 cm (17.24 inches) |
Dim. W. | 5.7 cm (2.24 inches) |