ETS-Lindgren HI-6153 Electric Field Probe 10 MHz – 40 GHz

Description

The laser powered HI-6153 Electric Field Probe provides broadband EMF frequency coverage and wide dynamic range that satisfies the demands of most test requirements. To take advantage of this capability, the HI-6153 was designed to be single range reading so data can be read continuously over the entire dynamic range. Data values for each axis (X, Y, and Z) can be read individually, or summed.

Fiber optic signal and power lines link the Model HI-6153 probe to either the model HI-6100 Field Monitor or as a direct connect to a PC with the HI-6113 Laser Data Interface and ProbeView Laser Software.

The HI-6100 provides manual functions and programmed control via IEEE-488 and RS-232 Serial Data Interfaces. Readings from up to four probes can be displayed simultaneously.

The HI-6113 Laser Data Interface provides the laser power and communications for the HI-6153 Electric Field Probe. A USB connection to the PC allows for quick and easy data collection, using ProbeView Laser software.

Features

  • Laser Powered for Unlimited Test Time
  • Frequency Range: 10 MHz – 40 GHz
  • Dynamic Range: 2.0 – 800 V/m (single range)
  • Provides Individual and Summed Axis Values
  • A2LA Accredited Calibration Report
  • Suitable for MIL Standard Specs:
    — MIL-STD 461F Radiated Susceptibility (RS)
  • Suitable for Automotive Specs:
    — SAE J1113/27
    — GM9120P
    — GM93100GS
    — GM9114P
  • Suitable for Commercial Specs: EN/IEC 61000-4-3 Radiated Immunities
Download Data Sheet
Print
Email

Specifications

Electrical Specifications
Detection Isotropic (X, Y, and Z Axis Readings)
Frequency Range 10 MHz to 40 GHz
Frequency Response with Correction 10 MHz to 18 GHz ± 0.9 dB
18 to 40 GHz ± 1.1 dB
Frequncy Response (Typical) 10 to 100 MHz + 3.0, -4.0 dB
100 MHz to 1 GHz +3.0, -0.50 dB
1 to 18 GHz +4.0, -2.0 dB
18 to 40 GHz +3.5, -4.5 dB
Dynamic Range 2 to 800 V/m (Single Range)
Resolution 0.01 V/m
Isotropic Deviation ±1.0 dB < 18 GHz
Linearity ±0.5 dB @ 1 GHz
Sample Rate (Typical) HI-4413P Serial Modem – 44 samples/second maximum

HI-6100 Field Monitor – 70 samples/second maximum

Overload Withstand >1500 V/m CW
Physical Specifications
Physical Interface Duplex Optical Fiber (62.5 micron multimode)
FC for laser, Integral 1m optical cable
Operating Temperature Range 10° to 40° C, 5-95% Relative Humidity Non-condensing
Dim. L. 43.8 cm   (17.24 inches)
Dim. W. 5.7 cm   (2.24 inches)
Weight 0.36 kg (12.64 oz)
Mounting 1/4-20 UNC Internal Thread