Haefely Hipotronics PEFT 4010 EFT/Burst Immunity Test System 4.8kV
Description
Haefely Hipotronics PEFT 4010 EFT/Burst Immunity Test System is used to measure the most common sources of disturbances in modern electronic circuits. EFT/bursts are caused by the operation of electro-mechanical switches, motors or power distribution switchgears and are generated according to the IEC/EN 61000-4-4.
The PEFT 4010 instrument contains all the features expected from an EFT generator. It can either be operated by the front panel keys in a standalone manner or be controlled by PC via RS232 or IEEE 488 as part of a complete EMC test system.
Features
- EFT impulse generation
- ESD control and report functions
- Automatic test operation
- Data exchange
- Compact and reliable construction
- Remote control
- ISO 9001:2000
Download Data Sheet
Print
Email
Specifications
Wave Shape into 50Ω load | |
Rise time | 5 ns ± 30 % |
Impulse duration | 50 ns ± 30 % |
Wave Shape into 1000Ω load | |
Rise time | 5 ns ± 30 % |
Impulse duration | 50 ns – 15 / + 100 ns |
Burst Characteristic | |
Voltage range | 0.22 … 4.8 kV
(0.1 … 4.8 kV in expanded mode) |
Spike frequency | 1 Hz … 1 MHz
(1 Hz … 2 MHz in expanded mode) |
Burst duration | 0.01 … 999 ms |
Burst period | 0.1 Hz … 400 Hz
2.5 ms … 10 s 1 period … 500 periods |
Impulses per second | Max. 12’000 |
Impulses per burst | Max. 1’000 |
Spike distribution | normal, continuous, random, real (real5, real100 with “Real
Burst” option) |
Polarity | positive, negative |
Test time | 10 s … 8 h per path |
Single Phase Coupling/Decoupling Network | |
Maximum AC voltage | 264 V @ 16 … 440 Hz |
Maximum AC current | 16 A @ 50 … 60 Hz, 10 A @ 400 Hz |
Maximum DC voltage | 125 V |
Maximum DC current | 16 A (dependant on DC voltage) |
Coupling Modes | L-GND, N-GND, PE-GND, LN-GND, LPE-GND, NPEGND, LNPE-GND |
Residual voltage at Test Supply Input | ≤ 10% of applied test voltage |
Synchronization | |
Frequency | 16 2/3 Hz, 40 Hz, 50 Hz, 60 Hz, 400 Hz, auto |
Impulse trigger | automatic, manual, external |
System Configuration | |
EUT fail input | BNC, logic low to trigger |
P90 interface | for a 3-phase CDN extension |
PESD interface | for connecting an ESD generator |
Printer interface | Centronix |
Computer interface | RS-232, IEEE-488 (optional) |
Power Supply | |
Voltage | 85 … 264 VAC, 150 VA |
Frequency | 50 / 60 Hz |
Weight and Dimension | |
Weight | 14 kg |
W x H x D | 450mm x 130mm x 570mm |