Anritsu MG3633A Synthesized Signal Generator 10 kHz - 2.7 GHz
Description
The MG3633A has excellent frequency resolution, frequency switching speed, signal purity, and a high output level in addition to amplitude, frequency, and phase modulation functions.
Also, sweep functions are provided for carrier frequency, output level, and modulation frequency so an appropriate sweep can be performed for various devices to be measured. Also, the MG3633A has a frequency memory that can store 1000 carrier frequencies and a function memory that stores 100 panel settings. Moreover, since the maximum output level is +17 dBm, it can be used for various local signal sources.
The MG3633A is suitable for research and development of mobile communications in the quasi-microwave band, performance evaluation, characteristics testing, and adjustment of various types of radio equipment such as digital land-based mobile communications, mobile satellite communications, satellite broadcasting, and radio LANs.
Features
Low Noise
Using both the latest synthesizer and RF-device technologies and optical data links in the internal control circuit, the SSB phase noise has been cut to -140 dBc/Hz (CW, 1.1 GHz, offset 20 kHz). In particular, the MG3633A shows its power in measurement of narrow-band radio equipment S/N ratio and adjacent channel selectivity.
High accuracy and high-output level
Low levels of -123 dBm can be set with ±1 dB accuracy by using a high-accuracy programmable attenuator. The output level can be displayed in units of dBm, dBµV, V, mV, and µV or as a relative value (dB).
Modulation characteristics
The MG3633A has AM, FM, oM, and a combination of all three modulation functions. A DC mode is provided for FM, which makes simulation of digital transmissions for a pager possible. Also, a built-in AF oscillator with a 0.1 Hz to 100 kHz synthesizer can handle various modulations.
Quasi-microwave output
The MG3633A covers a wide range (from 10 kHz to 2700 MHz) and is suitable for research and development, as well as production of quasi-microwave band radio equipment.
Specifications
Carrier frequency |
Range |
10 kHz to 2700 MHz |
||
Resolution |
0.01 Hz |
|||
Accuracy |
Same as that of the reference oscillator |
|||
Internal reference oscillator*1 |
Frequency: 10 MHz Start-up characteristics: After 30 minutes of operation: ≤1 x 10–7/day, after 60 minutes of operation: ≤5 x 10–8/day, Aging rate: After 24 hours of operation: ≤2 x 10–8/day, Temperature characteristics: ±5 x 10–8/day, (0˚ to 50˚C) |
|||
External reference signal input |
10 MHz, TTL Level, BNC connector on rear panel |
|||
Reference signal output |
10 MHz, TTL Level, BNC connector on rear panel |
|||
Switching time |
≤10 ms (time from last command until frequency has stabilized to within ±500 Hz of set frequency, during remote operation) |
|||
Output |
Range |
–143 to +23 dBm |
||
Units |
dBm, dBµV, V, mV, µV (Terminated and open voltages are selectable for dBµV, V, mV or µV.) |
|||
Resolution |
0.1 dB |
|||
Frequency response |
±0.5 dB referred to 0 dBm (<1280 MHz), ±1 dB referred to 0 dBm (≥1280 MHz) |
|||
Accuracy |
Frequency Output level |
10 kHz to <1280 MHz |
≥1280 MHz |
|
+17.1 to +23 dBm |
– |
– |
||
+15.1 to +17 dBm |
±1 dB |
– |
||
–122.9 to +15 dBm |
±1 dB |
±2 dB |
||
–132.9 to –123 dBm |
±3 dB |
±4 dB |
||
–143 to –133 dBm |
– |
– |
||
Impedance |
50 Ω, N-type connector VSWR: ≤1.5 (<1280 MHz, ≤–3 dBm), ≤1.8 (≥1280 MHz, ≤–3 dBm) |
|||
Switching time |
Time from last command until output level is stabilized, during remote operation:≤25 ms (at LEVEL NORMAL mode) ≤80 ms (when setting level is crossing over –59 dBm, at LEVEL NORMAL mode) ≤5 ms (at LEVEL CONTINUOUS mode) |
|||
Interference radiation |
≤1 µV (Value is voltage terminated with 50 Ω load, measured 25 mm from front panel with a two-turn 25 mm diameter loop antenna.) Except sweep mode |
|||
Range |
10 kHz to 2700 MHz |
|||
Resolution |
0.01 Hz |
|||
Accuracy |
Same as that of the reference oscillator |
|||
Internal reference oscillator*1 |
Frequency: 10 MHz Start-up characteristics: After 30 minutes of operation: ≤1 x 10–7/day, after 60 minutes of operation: ≤5 x 10–8/day, Aging rate: After 24 hours of operation: ≤2 x 10–8/day, Temperature characteristics: ±5 x 10–8/day, (0˚ to 50˚C) |
|||
External reference signal input |
10 MHz, TTL Level, BNC connector on rear panel |
|||
Reference signal output |
10 MHz, TTL Level, BNC connector on rear panel |
|||
Switching time |
≤10 ms (time from last command until frequency has stabilized to within ±500 Hz of set frequency, during remote operation) |
|||
Range |
–143 to +23 dBm |
|||
Units |
dBm, dBµV, V, mV, µV (Terminated and open voltages are selectable for dBµV, V, mV or µV.) |
|||
Resolution |
0.1 dB |
|||
Frequency response |
±0.5 dB referred to 0 dBm (<1280 MHz), ±1 dB referred to 0 dBm (≥1280 MHz) |
|||
Accuracy |
Frequency Output level |
10 kHz to <1280 MHz |
≥1280 MHz |
|
+17.1 to +23 dBm |
– |
– |
||
+15.1 to +17 dBm |
±1 dB |
– |
||
–122.9 to +15 dBm |
±1 dB |
±2 dB |
||
–132.9 to –123 dBm |
±3 dB |
±4 dB |
||
–143 to –133 dBm |
– |
– |
||
Impedance |
50 Ω, N-type connector VSWR: ≤1.5 (<1280 MHz, ≤–3 dBm), ≤1.8 (≥1280 MHz, ≤–3 dBm) |
|||
Switching time |
Time from last command until output level is stabilized, during remote operation:≤25 ms (at LEVEL NORMAL mode) ≤80 ms (when setting level is crossing over –59 dBm, at LEVEL NORMAL mode) ≤5 ms (at LEVEL CONTINUOUS mode) |
|||
Interference radiation |
≤1 µV (Value is voltage terminated with 50 Ω load, measured 25 mm from front panel with a two-turn 25 mm diameter loop antenna.) Except sweep mode |
|||
External modulation |
Input level: Approx. 2 Vp-p/600 Ω Input impedance: Nominal 600 Ω |
|||
Distortion |
≤1% (internal 1 kHz, 5 rad modulation) |
|||
400 Hz, 1 kHz (fixed oscillator) 0.1 Hz to 100 kHz (variable oscillator) DC voltage signals equivalent peak values of internal modulating sine wave can be applied as a modulating signal using the SPECIAL FUNCTION. |
||||
Resolution |
0.1 Hz |
|||
Frequency accuracy |
100 ppm |
|||
Distortion |
≤0.03% (fixed, 400 Hz and 1 kHz), ≤0.3% (variable, 20 Hz to 50 kHz) |
|||
Frequency memory |
1000 carrier frequencies (store/recall) |
|||
Function memory |
100 panel settings (store recall) |
|||
Sweep mode |
Carrier frequency, output level, AF frequency |
|||
Sweep time |
0.1 ms to 600 s, 0.01 ms resolution (minimum time depends on the switching time of each function.) |
|||
Marker |
One movable marker |
|||
Sweep signal output |
Staircase (saw-tooth waveform), Start point: 0 V, Stop point: 10 V |
|||
Modulation signal output |
Modulation signal is output when modulating. Output level: Approx. 2 Vp-p/600 Ω |
|||
Simultaneous modulation |
Simultaneous modulation is possible in combinations shown below. |
|||
Relative value display |
Carrier frequency, output level |
|||
Continuously variable output level mode |
Continuously variable within a ±10 dB range of the set level Step size: 0.1 dB |
|||
Trigger function |
Previously programmed operation procedure can be started by a trigger input through its input terminal (on rear panel, BNC connector, TTL level). Maximum program steps for triggered operation: 99 steps |
|||
Memory backup |
Last settings are stored when power is turned off. |
|||
GPIB |
Interface function: SH1, AH1, T5, L3, TE0, LE0, SR1, RL1, PP0, DC1, DT1, C0 |
|||
Reverse power |
Maximum reverse input power: 50 W (<1000 MHz), 25 W (≥1000 MHz), ±DC 50 V |
|||
Operating temperature |
0˚ to 50˚C |
|||
Power |
*7Vac +10%, 48 to 63 Hz, ≤270 VA –15 |
|||
Dimensions and mass |
426 (W) x 177 (H) x 451 (D) mm, ≤32 kg |
|||
EMC*8 |
EN55011: 1991, Group 1, Class A EN50082-1: 1992 |
|||
Safety |
EN61010-1:1993 (Installation Category II (Pollution Degree II) |