Amplifier Research FA7000 Electric Field Analyzer 100 kHz – 60 GHz
Description
Amplifier Research FA7000 series of Field Analyzers represent a new patent-pending approach to more accurately measure modulated electric fields as well as CW electric fields. This innovative approach uses an isotropic field sensor to sample the composite field and transmit its amplitude digitally over optical fiber to a processor unit. The sample rate of the FA7000 Field Analyzer is 1 .5 million samples per second – significantly faster than conventional RF field probes – enabling them to accurately measure pulsed electric fields in the microsecond range.
Each of the FA7000 series analyzer kits provides a web based oscilloscope-type display of the instantaneous electric field strength (V/m over time) and calculates the minimum, maximum, and average field strength of the waveform as displayed. Each kit consists of an isotropic field sensor, glass fiber-optic cabling, and a processor unit. The processor unit stores all of the necessary amplitude corrections for its associated field sensor.
Correction factors for many frequencies are provided with each kit. These factors can be loaded into the processor unit to automatically correct the field readings at user specified frequencies. When correction factors are applied, the true accuracy of this device is realized.
Features
- Faster, easier and more accurate
- Significantly faster than conventional RF field probes
- Measures modulated electric fields as well as CW electric fields
- 1.5 million Samples per second
- Includes isotropic field sensor, glass fiber-optic cabling and a processor unit
- Correction factors included
- Remote interfaces include LAN, USB 2.0 & IEEE-488
Specifications
Processor Unit |
|
Dimensions (W x H x D) |
21.91 x 4.45 x 27.69cm |
Field Processor Weight |
1.36kg |
Operating Temperature Range |
10˚C to 40˚C @ 5% to 95% RH non-condensing |
Fiber Optic Connector |
E2000 compact duplex (Yellow, keying #3) |
Fiber Optic Cable length |
20m (supplied with kit) |
Max Fiber Optic length |
100m (sold separately) |
Remote Interfaces |
LAN (Ethernet) |
USB 2.0 (Test and Measurement class) |
|
IEEE-488 (GPIB) |
|
Fiber Optic Serial (FSMA connectors; Reserved for use with FM7004 Field Monitor) |
|
Max Remote Transfer Rate |
20 queries per second |
External Trigger Port |
|
Impedance |
>10MΩ |
Threshold Voltage |
3V |
Maximum Input Voltage |
5V |
Minimum Pulse Width |
40ns |
Readout Display |
Embedded Web Application through PC (PC not included) |
Remote Interface |
LAN (Ethernet) |
Compatible Web Browsers |
Chrome, Internet Explorer, Safari, Firefox, Opera |
Time base Range |
1μs/Div to 400μs/Div |
Scale Range |
0.1(V/m)/Div to 500(V/m)/Div |
Trigger Modes |
Free Run, Internal (conventional threshold), External |
Edges (Threshold trigger) |
Rising and Falling |
Vertical Divisions |
10 |
Horizontal Divisions |
10 |
Laser |
|
Wavelength |
850nm |
Output Power |
≤ 500mW |
Shutdown Time |
< 1 ms after fiber disconnect |
Power Requirements |
|
Input Voltage |
90–260VAC, 50-60Hz |
Input Current |
0.2–0.6A |
Input type |
IEC inlet with filter |
Sample Rate |
1.5MS/s |
Max Record Length |
6kPoints |
Modulation Frequency Range |
250Hz to 750kHz |
Measurement Format |
Composite only |
SPECIFICATIONS FA7000 Series
Field Sensors |
FA7006 |
FA7218 |
FA7040 |
FA7060 |
Amplitude Accuracy1 |
±1.0dB @ 10MHz1 0.8dB, 100kHz–1GHz3,4 1.4dB, 1GHz–6GHz3,4 |
±1.0dB @ 10MHz2 0.8dB, 2MHz–1GHz3,4 1.4dB, 1GHz–18GHz3,4 |
±1.0dB @ 10MHz2 0.8dB, 2MHz–1GHz3,4 1.4dB, 1GHz–40GHz3,4 |
±1.0dB @ 10MHz2 0.95dB, 2MHz–1GHz3,4 1.5dB, 1GHz–60GHz3,4 |
Isotropic Deviation4,5 |
±1.2dB @ 10MHz ≤ 200V/m |
±1.0dB @ 10MHz ≤ 200V/m |
±1.0dB @ 10MHz ≤ 200V/m |
±1.0dB @ 10MHz ≤ 200V/m |
±2.0dB @ 10MHz >200V/m |
±2.0dB @ 10MHz > 200V/m |
±2.0dB @ 10MHz > 200V/m |
±2.0dB @ 10MHz > 200V/m |
|
Operating Range9 |
9 – 900V/m |
14–1400V/m |
14–1400V/m |
14–1400V/m |
Linearity |
±0.5dB |
±0.5dB |
±0.5dB |
±0.5dB |
Typical Analog Rise Time6,7 |
300nS |
300nS |
300nS |
300nS |
Minimum Pulse Width |
2μs |
2μs |
2μs |
2μs |
Damage Level (CW) |
1000V/m |
1200V/m |
1200V/m |
1200V/m |
Temperature Stability |
±1.0dB, 10°C – 40°C8 |
±1.0dB, 10°C – 40°C8 |
±1.0dB, 10°C – 40°C8 |
±1.0dB, 10°C – 40°C8 |
Approximate Dimensions (W x H x D) |
5.7 x 5.7 x 5.7cm |
27.8 x 6.5 x 6.5cm |
27.8 x 6.5 x 6.5cm |
27.8 x 6.5 x 6.5cm |
Weight |
62.5g |
150g |
150g |
150g |
1 Single axis aligned with field
2 Without correction factors applied
3 With correction factors applied
4 Typical expanded measurement uncertainty (95% confidence interval)
5 Measured at the ortho angle
6 10% – 90%
7 Pre-digitization
8 5% – 95% Relative humidity, non-condensing
9 Less than 50% duty