Amplifier Research MT06000 RF Radiated Immunity System 80MHz – 6GHz

Description

Amplifier Research MT06000 (Multistar™ Multi-tone tester) is a state-of-the-art system designed to test RF Radiated Immunity faster than ever before possible. By testing multiple frequencies (tones) at once, test times can be reduced by a factor equivalent to the number of tones selected. The number of tones is only limited by the signal generator bandwidth (150 MHz) and the size of the amplifier used with the system.

The MT06000 contains all the instruments needed to perform radiated immunity testing for IEC 61000-4-3 specification except the required amplifiers, antennas and directional couplers. Amplifiers can be sized and selected based on your required field levels and testing needs.

Up to 4 RF amplifiers and directional couplers can be controlled and monitored and power can be delivered to up to 4 antennas to generate the desired fields. The system contains a vector signal generator, a vector signal analyzer, a RF pre-amplifier, a RF field probe and monitor, an RF switch matrix, and automated radiated immunity test software.

Everything is contained in a single housing, which eliminates setup issues. The software includes automated routines to calibrate the field and maximize the speed of test (most tones possible) while still meeting the Linearity and Harmonics requirements of the specification.

In the event of a EUT failure, margin investigation (thresholding) and traditional single tone testing can be performed causing a slowing of the test only in the areas of concern. This system has the versatility needed for every test laboratory and equipment manufacturer while adding the benefit of reduced test times and greater throughput.

 

Features

Complete Testing Solutions to the following standards

  • EN/IEC 61000-4-3 
  • EN/IEC 60601-1-2 
  • EN 50130-4 
  • EN 61000-6-1/2 
  • EN 55024
Download Data Sheet
Print
Email

Specifications

Internal Test Specifications

IEC/EN 60601-1-2

IEC/EN 50130-4

IEC/EN 61326

IEC/EN 61000-6-1

IEC/EN 61000-6-2

CISPR 24/EN 55024

IEC 61000-4-3 procedure and levels

 

Vector Signal Generator Specifications

Frequency range Resolution

80 < Hz to 6.6 GHz < 4 Hz

Power Out

+10 dBm

Modulation

AM, FM, Pulse, Phase

Bandwidth

150 MHz

Hardware Platform

PXIe

 

Vector Signal Analyzer Specifications

Frequency Range Resolution

10 MHz – 6.6 GHz 533 nHz

Dynamic Range

80 dB

Bandwidth

50 MHz

Hardware Platform

PXIe

 

RF Preamplifier Specifications

Frequency Range

50 MHz – 8 GHz

Gain

26 dB

Hardware Platform

PXIe

 

Field Monitor/Probe Spec

Channels

4

Probe

1

Type

Isotropic, Laser powered

Frequency

100 kHz – 6 GHz

Range

0.5 – 800 V/m

 

Connections

RF signal input

2 SMA Male (rear)

For optional signal generators

RF Signal Out

4 SMA Male (rear) to RF amplifiers

High power RF in

4- Type N Male (rear) from RF amplifiers

High power RF out

4 Type N male (rear) to antennas/loads

Monitor port In

4- SMA Male (rear)

4 for forward power

 

Control PC

Computer

Intel Core 2 Duo, 2.53 GHz

Operating system

Windows 7

RAM

4 GB DDR2

Hardware Platform

PXIe

 

General

Power

115/230 VAC

50/60 Hz, single phase 16A

Breaker

2 pole, 20A

Cooling

active cooling, air ventilation

Environmental conditions

10°C – 40°C

Dimensions

50.3 x 47.2 x 61 cm, 19.8 x 18.6 x 24 in

Weight

22.7 kg (50.0 lb)

 

Accessories

DAQ Module (AR P/N 10025318) (NIUSB-6212)

Data acquisition module is used to collect data, identify user defined failures, report user defined error messages, and reset EUT after failure

Low Pass Filter, 80MHz-1GHz

Absorptive filter used to remove harmonics

Low Pass Filter, 80MHz-4.2GHz

Absorptive filter used to remove harmonics

Low Pass Filter, 80MHz-6GHz

Absorptive filter used to remove harmonics

 