Anritsu S331A Cable and Antenna Analyzer 25 MHz - 3.3 GHz
Description
The Site Master S331A is a hand held SWR/RL (standing wave ratio/return loss), and Distance-To- Fault (DTF) measurement instrument that includes a built-in synthesized signal source. All models include a keypad to enter data and a liquid crystal display (LCD) to provide graphic indications of SWR or RL over the selected frequency range and selected distance.
The Site Master is designed for measuring SWR, return loss, or cable insertion loss and locating faulty RF components in antenna systems. Power monitoring is available as an option. The displayed trace can be scaled or enhanced with frequency markers or a limit line. A menu option provides for an audible “beep” when the limit value is exceeded. To permit use in low-light environments, the LCD can be back lit using a front panel key.
The Site Master Software Tools program also performs DTF (Distance to Fault) and Fault Location.
Features
- Accurate Test of Antenna Systems and Active/Passive Components
- Built-in Distance-To-Fault
- Immunity to Live Site RF Interference
- Synthesizer Accurate to 75 ppm
- Trace Storage, 40 Max.
- Automatically date/ time stamp
- Alphanumeric labeling give distinct names
- Instrument calibration is valid in all data point setting
- 10 user configurations include calibrations
- Silica Glass Based Optical Fibers
- Math Functions: Trace – Memory, Trace + Memory
- Direct Printer Control via RS232 Serial Port
Specifications
Frequency Range |
25 to 3300 MHz |
|
Frequency Accuracy |
(RF Source Mode) 75 parts per million @ 25°C* |
|
Frequency Resolution |
100 kHz |
|
SWR: |
Range |
1.00 to 65.00 |
|
Resolution |
0.01 |
Return Loss |
Range |
0.00 to 54.00 dB |
|
Resolution |
0.01 db |
Test Port, Type N 50 Ohms ***Immunity to Interfering signals up to the level of –5 dBm, reflection |
||
Maximum Input (Damage Level) |
Test Port, Type N: +22 dBm |
|
RF Detector: |
+20 dBm |
|
Measurement Accuracy |
Measurement accuracy depends on calibration components |
|
Precision calibration components have a directivity of 42 dB |
||
Cable Insertion Loss |
Range |
0.00 to 54.00 dB |
Resolution |
0.01 dB |
|
Frequency Reference |
Aging |
±1 ppm/yr |
Accuracy |
±2 ppm |
|
Internal Memory: |
Trace Memory |
40 maximum |
Instrument configuration |
10 setup locations |
|
RS-232: 9 pin D-sub, three wire serial |
||
Electromagnetic Compatibility Meets European community requirements for CE marking |
||
Power Supply |
External DC Input: +11 to +15 Vdc, 1250 mA max |
|
Temperature |
Storage: –20° C to 75° C |
|
Operation: 0° C to 50° C |
||
Weight |
4 lbs (1.82 kg) |
|
Dimensions |
10 x 7 x 2.4 in (25.4 x 17.8 x 6.1 cm) |