Fluke PM6304 RCL Meter

Description

If you want an instrument with remote programmability, consider the Fluke PM 6304 RCL meter. The PM 6304 has a basic accuracy of 0.1%, 3 test voltage levels and 204 test frequencies. The PM 6304C has an even higher basic accuracy, and can measure components or devices with an accuracy of 0.05%. Thanks to this high accuracy, the PM 6304C is ideally suited for calibrating the capacitance sourcing function of the high-performance Fluke Calibration 5500A Multi-Product Calibrator.

 

Features

  • Select the test you need
  • Easy-to-use, at-a-glance display
  • Test frequencies from 50 Hz to 1 MHz (100 kHz for PM 6304)
  • 0.1% basic accuracy (0.05% for PM 6304C)
  • RS-232 or IEEE-488 interface
  • AC test levels from 50 mV to 2V rms
  • Internal or external bias
  • DC test measurement (optional)
  • 9 front panel setups
  • Actual component test voltage/current readback
  • Automatic zero trimming
  • Contact check and deviation mode (PM 6306 only)
  • Front-panel test posts for immediate 4-wire measurements
Download Data Sheet
Print
Email

Specifications

AC Test Mode

Test frequency

PM 6306:

50, 60, 100, 125 Hz
250 Hz to 100 kHz (100 Hz steps)
100 kHz to 1 MHz (1 kHz steps)

PM 6304:

50, 60, 100, 125 Hz
250 Hz to 25 kHz (100 Hz steps)
100 kHz

Test frequency accuracy

PM 6306:

0.01%

PM 6304:

0.01%

Test signal levels

PM 6306:

50 mV to 2V (10 mV steps) via 100Ω

PM 6304:

50 mV via 100Ω
1V via 100Ω
2V via 400Ω

Basic measurement accuracy at normal measurement mode

PM 6306:

0.1% ± 1 digit (for < 0.25V, > 50 kHz)
0.1% ± (f/50 kHz) ± 1 digit (for < 0.25V, > 50 kHz)
0.1% ± (0.25V/Vt) ± 1 digit (for < 0.25V, > 50 kHz)

PM 6304:

0.05% ± 1 digit (for PM 6304C, > 2 kHz)
0.1% ± 1 digit (for > 25 kHz)
0.4% ± 1 digit (100 kHz)
0.5% ± 1 digit (for 50 mV > 25 kHz)
2.0% ± 1 digit (for 50 mV, 100 kHz)

DC bias

Internal

PM 6306:

0 to 10V (0.1V steps)

PM 6304:

2V

External

PM 6306:

0 to 40V

PM 6304:

0 to 40V

DC test mode (Optional)

Test signal levels

PM 6306:

50 mV to 2V (10 mV steps) via 100Ω

PM 6304:

300 mV via 100Ω
1V via 100Ω
2V via 400Ω

Basic measurement accuracy

PM 6306:

0.1% ± 1 digit (for < 0.25V)

PM 6304:

0.1% ± 1 digit (for 1V)

Contact check (PM 6306 only)

Pass

3 3Ω

Fail

3Ω (with indication of failed connection lead)

Maximum measuring ranges

Impedance/Resistance AC Z or RAC

0.0000O to 250 MΩ

Resistance DC R DC

0.0000O to 50 MΩ

Capacitance C

0.00 pF to 31.8F

Inductance L

0.00 mH to 637 kH

Quality factor Q

0.000 to 1000

Dissipation factor D

0.000 to 1000

Phase Angle F

-179 to + 175 deg

Voltage monitor VX

0.1 mV to 2.00V

Current monitor IX

0.005 mA to 10.0 mA

Maximum resolution

Impedance/Resistance AC Z or RAC

0.1 mΩ

Resistance DC R DC

0.1 mΩ

Capacitance C

0.01 pF

Inductance L

0.01 uH

Quality factor Q

0.001

Dissipation factor D

0.001

Phase Angle F

0.1 deg

Voltage monitor VX

0.1 mV

Current monitor IX

0.001 mA

Circuit diagram

Display

1 of 7 different equivalent circuit diagrams

Auto Mode

Read-out

Dominant and secondary parameter

Equivalent circuit diagram

Parallel for R+C, Serial for R+L

Manual mode

Read-out

Dominant and secondary parameter or Z, F, D, Q, VX, IX

Equivalent circuit diagram

Parallel or serial selectable

Average function

Function

Exponential averaging in continuous mode

Levels

PM 6306:

3 (and off)

PM 6304:

1 (and off)

Deviation mode (PM 6306 only)

Relative range in respect to reference value

-100% to +100%

Measuring modes (Normal)

Continuous

2 measurements/s

Single

Triggered via “TRIG” key
Triggered via handler interface
Triggered via IEEE-488 or RS-232

Test frequency

PM 6306:

50, 60, 100, 125 Hz
250 Hz to 100 kHz (100 Hz steps)
100 kHz to 1 MHz (1 kHz steps)
DC (optional)

PM 6304:

50, 60, 100, 125 Hz
250 Hz to 25 kHz (100 Hz steps)
100 kHz
DC (optional)

Read-out

Display or via IEEE-488 or RS-232 interface

Measuring modes (Fast)

Max. speed

10 measurements/s

Test frequency

PM 6306:

250 Hz to 100 kHz (250 Hz steps)
100 kHz to 1 MHz (1 kHz steps)
DC (optional)

PM 6304:

250 Hz to 25 kHz (250 Hz steps) 100 kHz

 

DC (optional)

Single

Triggered via handler interface
Triggered via IEEE-488 or RS-232

Read-out

Via IEEE or RS-232 interface (display blanked)

Binning

Standard bins

9

Special bins

Bin “0” and bin “fail”

Bin programming via

IEEE-488 interface

 

RS-232 interface

 

Bin programmer (PM 6304 only)

Bin limit programming

Absolute or relative

Trim function

Open circuit

Open circuit compensation Z > 100 kW

Short circuit

Short circuit compensation Z < 10W

Protection against charged capacitors

C < 2 mF

PM 6306:

Vmax < 250V

PM 6304:

Vmax < 500V

2 mF ≤ C ≤ 2 mF

PM 6306:

Vmax <47 x (C/mF) -0.234

 

(Vmax in V, C in mF)

PM 6304:

Vmax <117 x (C/mF) -0.234

 

(Vmax in V, C in mF)

Stored settings (non-volatile memory)

Front panel settings

9 + 1 (trim figures included)

Bin settings

9 + 1

Print measurements results

Print measurements results

Via RS-232 interface for serial printers

Calibration interval

Calibration interval

1 year

Environmental Conditions

Operating Temperature

0° to 50° C

Storage Temperature

-40° C to 70° C

Power, battery life

100/125/225/240V + 10%

Line frequency

50/60 Hz

PM 6306:

44 VA

PM 6304:

31 VA

Power Consumption

 

Safety

According to CE-regulation 73/23 EN61010-1 CAT II, Pollution Degree 2

EMC

According to CE regulation 89/336: Emission according to EN 55011, Group 1 Class B, respectively CISPR 11. Immunity according to EN 50082-1, inclusive IEC 751-2, -3, -4

Warm up Time

PM 6306:

30 minutes

PM 6304:

5 minutes

Dimensions and weight

W x H x D

315 x 105 x 405 mm (12.4 in x 4.13 in x 15.9 in)

Weight

PM 6306:

5.3 kg (11.7 lb)

PM 6304:

4.7 kg (10.4 lb)

Stay informed