Solar 2854-1 Transient Generator

Description

For MIL-STD-461F CS106 Conducted Transient Susceptibility Test.

The generator provides 5 μS ±22 %, pulse with a 1.5 μS ± 0.5 μS rise time across a 5.0 ohm non-inductive resistor. Voltage is adjustable to greater than 600 volts. The undershoot is limited to less than 120 volts peak (maximum) and less than 20 μS. The repetition rate is variable and can be adjusted from 0.8 p.p.s. to 10 p.p.s. Single transients can be applied with the pushbutton on the front panel.

Two sets of output terminals allow either parallel or series injection into the power line. The Series injection may be used on D.C. and A.C. lines. Parallel injection should be used on D.C. lines only. The output winding used for series injection can carry 25 amperes of power current. The output terminals are isolated from the chassis and the power cord.

Features

  • Provides output levels from 10 volts to more than 600 volts into 5 ohms
  • Adjustable pulse position on A.C. lines relates the transient susceptibility to real time aspects of digital systems
  • Single pulse feature for controlled isolation of transient effects
  • Output terminals for series and parallel injection
  • Standard rack panel construction: 7” high, 19” wide, 12.75” deep (17.78 cm wide, 48.26 cm high, 32.38 cm deep)

Solar 2854-1 Application

Solar Model 2854-1 Transient Generator is designed for screen room use in making conducted transient susceptibility tests as specified in MIL-STD-461F, Test Method CS106.

Download Data Sheet
Print
Email

Specifications

Spike Amplitude Continuously adjustable from 10 volts to over 600 volts peak
Repetition Rate Continuously adjustable from 0.8 to 10 p.p.s
Rise time 1.5 ± 0.5 μS into 5 ohms
Spike Duration Output falls to zero in 5 μS ± 22 %
Spike Shape Ringing characteristic as shown in Figure CS106-1 in MIL-STD-461F
Phase Adjustment position adjustable from 0° to 360° periodically on 50, 60 or 400 Hz sine wave
Internal Impedance less than 2.0 ohms
Power Requirements 115 volts 60 Hz, 1.6 amperes (230 volts, 50 Hz, 0.8 amperes available)
Size Standard rack panel 7” high x 19” wide x 12.75” deep (17.78 cm x 48.26 cm x 32.38 cm)