Solar 8282-1 Transient Pulse Generator

Description

For Conducted Transient Susceptibility Testing, 0.15 µS – 5.0 µS – 10.0 µS up to 600 volts (peak)

The Solar 8282-1 Transient Pulse Generator was designed for screen room use in making conducted spike susceptibility tests. It provides all the wave shapes required by MIL-STD-461B/C and many other military EMI specifications.

Spike generators required for susceptibility testing have been our specialty since 1962, when our first unit, the Model 6254-1, was introduced. The Model 8282-1incorporates all the flexibility and technical excellence of the previous models and provides features required by specification MIL-STD-461B/C.

 

Features

  • Provides outputs up to 600 volts peak amplitude for the 0.15, 5.0 and 10.0 µS spikes into a five ohm resistive load (low source impedance).
  • A wide range of repetition rates allows spike injection in terms of the pulse rates of items being tested.
  • The single pulse feature enables controlled isolation of transient effects.
  • Adjustable pulse position on AC power lines relates the transient susceptibility to the real time aspects of digital circuitry served by AC power.
  • Transients can be injected in synchronism with repetitive circuit functions as required by Method CS06 of MIL-STD-462.
  • Remote triggering of single or repetitive pulses in terms of particular system characteristics.
  • The upper terminals of the PARALLEL pair and the SERIES pair provide a positive-going spike on the 5 µS and the 10 µS modes. These terminals deliver a negative-going spike in the 0.15 µS mode. When the test plan requires both a positive and a negative spike, it is necessary to reverse the connections to the output terminals of the Model 8282-1 Spike Generator.
Download Data Sheet
Print
Email

Specifications

Spike Durations

Pushbutton selectable durations of 0.15 µS, 5.0 µS and 10.0 µS (±20%) to zero crossover, into 5.0 ohm resistive load

Adjustable Peak Amplitude

Up to 600 volts for 0.15 µS, 5.0 µS and 10.0 µS durations into five ohm non-inductive load

Internal Impedance

Less than 5.0 ohms for 0.15 µS, less than 2.0 ohms for 5.0 µS, less than 1.0 ohm for 10.0 µS

Pulse Repetition Rate

Less than 5.0 ohms for 0.15 µS, less than 2.0 ohms for 5.0 µS, less than 1.0 ohm for 10.0 µS

Pulse Shape

Ringing characteristic similar to Figure 19 in MIL-STD-462 when connected to non-inductive load

Pulse Position

Adjustable from 0° to 360° on 50 Hz, 60 Hz or 400 Hz power lines

External Sync Operation

Remotely trigger able up to 50 p.p.s. for 0.15 µS, up to 1000 p.p.s. for 5.0 and 10.0 µS

Amplitude Display

Panel meter is analog LED display of peak amplitude as it would be into a five ohm resistive load

Power Current in Series Injection Mode

up to 50 amperes of current at power frequencies

Power Requirements

115 volts 60 Hz, 3.0 amperes (230 volts 50 Hz, 1.5 ampere available)

Size

12.25″ wide , 8.7″ high 13″ deep (311 mm x 211 mm x 330 mm)

Weight

30 pounds