Description

The MT8820A covers the frequency range from 30 MHz to 2.7 GHz and is designed to perform Transmitter and Receiver tests on 3G W-CDMA UE terminals in accordance with 3GPP TS34.121 and GSM terminals for both R&D and manufacturing applications. With appropriate measurement software and hardware options installed, this single platform supports evaluation of all the main transmission/reception test items for W-CDMA terminals including HSDPA. Options also support EDGE/EGPRS and 1xEV-DO, as well measurements for two phones simultaneously.

3GPP call processing is also supported, allowing both functional tests and transceiver measurements to be made in a call processing state. In addition, the voice signal from the terminal can be echoed-back during conversation in order to perform a simple voice communications test. Audio codecs are also available for real-time communication with terminals.

Advanced DSP and parallel measurement technologies dramatically reduce wireless manufacturing and inspection test times. Furthermore, multiple measurements can be selected for batch measurement. A one-touch operation allows selected batch measurements to be executed repeatedly a designated number of times. Pass/fail evaluation of the main measurement items — including transmission frequency, modulation accuracy, output power, adjacent channel power, occupied frequency bandwidth, and BER — can be performed easily and quickly.

 

Features

  • Frequency range 30 MHz to 2.7GHz
  • Advanced DSP and parallel measurement technologies for fast test time
  • 3GPP call processing including Registration; Origination, and Termination
  • 12 Transmitter measurements items including Modulation analysis; Adjacent channel power ratio and Peak code domain error according to 3GPP TS34.121
  • Bit Error rate (BER) and Block Error rate (BLER) receiver measurements using 3GPP loopback method
  • “One-touch” Batch measurement operating mode with Pass/Fail evaluation
  • Multi-system W-CDMA / GSM capability
  • 1xEV-DO support including external packet data communications measurement capability and IP data communications mode
  • EDGE/EGPRS measurement software supporting MCS1-MCS4 (GMSK modulation) and MCS5-MCS9 (8PSK modulation) coding
  • HSDPA call processing and RF transmission/reception measurements for mobile terminals with UE category 1 to 6, 11, and 12
  • Parallel phone measurement to increase measurement speed, reduces production costs, and save manufacturing space

 

Download Data Sheet
Print
Email

Specifications

General

Frequency range: 30 to 2700 MHz

Max. input level: +35 dBm Main 1)

Main 1 input/output connector

Impedance: 50 Ω

VSWR: ≤ 1.2 (< 1.6 GHz), ≤ 1.25 (1.6 to 2.2 GHz), ≤ 1.3 (> 2.2 GHz)

Connector: N type

AUX 1 output connector

Impedance: 50 Ω

VSWR: ≤ 1.3 (at SG Output level: ≤ –10 dBm)

Connector: SMA type

Reference oscillator

Frequency: 10 MHz

Level: TTL

Startup characteristics: ≤ ±5 x 10–8 (at 10 min after startup referenced to frequency 24 h after startup)

Aging rate: ≤ ±2 x 10–8/day, ≤ ±1 x 10–7/year (referenced to frequency 24 h after startup)

Temperature characteristics: ≤ ±5 x 10–8

Connector: BNC type

External reference input

Frequency: 10 MHz or 13 MHz (±1 ppm)

Level: ≥ 0 dBm

Impedance: 50 Ω

Connector: BNC type

RF signal generator

Frequency

Frequency range: 30 to 2700 MHz (setting range: 0.4 to 2700 MHz)

Setting resolution: 1 Hz

Accuracy: Due to reference oscillator accuracy

Output level

Level range: –140 to –10 dBm (Main 1), –130 to 0 dBm (AUX 1)

Resolution: 0.1 dB

Accuracy: ±1.0 dB (–120 to –10 dBm, Main 1, after calibration), ±1.0 dB (–110 to 0 dBm, AUX 1, after calibration)

Signal purity

Non-harmonic spurious:

≤ –50 dBc (at offset frequency: ≥ 100 kHz, except Uplink frequency – Downlink frequency + 4.1825 GHz),

≤ –40 dBc [spurious of (4.8 GHz – Downlink frequency GHz) GHz at ≥ 2.1 GHz]

Harmonics: ≤ –25 dBc

Uninterrupted level variation

Variable range: 0 to –30 dB

Setting resolution: 1 dB

Others

Display

Color 8.4″ TFT LCD, 640 x 480 dots

External control

GPIB: Control from external host with main unit as device (excluding some functions such as power-on),

no external device control

Interface functions: SH1, AH1, T6, L4, SR1, RL1, PP0, DC1, DT1, C0, E2

Power supply

100 to 120/200 to 240 Vac (–15/+15%, 250 V max.), 47.5 to 63 Hz, ≤300 VA (with Option 01), ≤650 VA (with all Options)

Dimensions and mass

426 (W) x 221.5 (H) x 498 (D) mm (excluding projections), ≤ 27 kg (with Option 01), ≤ 34 kg (with all Options)

Environmental conditions

0 to +50°C, ≤ 95% (no condensation)

Storage temperature and humidity: –20 to +60°C, ≤ 95% (no condensation)

EMC

EN61326: 1997/A2: 2001 (Class A), EN61000-3-2: 2000 (Class A), EN61326: 1997/A2: 2001 (Annex A)

LVD

EN61010-1: 2001 (Pollution Degree 2)

 

System

Software

Model

W-CDMA

W-CDMA Measurement Software

MX882000B

W-CDMA Call Processing Software

MX882050A

HSDPA Measurement Software

MX882200B-11

W-CDMA Ciphering Software

MX882071A

GSM/GPRS

GSM Measurement Software

MX882001A

EGPRS

EGPRS Measurement Software

MX882001A-11

cdma2000 1x

cdma2000 Measurement Software

MX882002A

cdma2000 Wireless Application Test Software

MX882022A

cdma2000 1xEV-DO

1xEV-DO Measurement Software

MX882003A

PDC

PDC Measurement Software

MX882004A

PHS

PHS Measurement Software

MX882005A