Anritsu S332D Cable & Antenna / Spectrum Analyzer 25 MHz - 4 GHz

Description

Site Master – Handheld Cable and Antenna Analyzer Model S332D includes all the features of our popular S332D – PLUS spectrum analysis for extra versatility in a single instrument. The standard cable & antenna analyzer spans from 25 MHz to 4 GHz. Option 2 and Option 16 extend the lower and upper frequency ranges of the cable & antenna analyzer to 2 MHz and 6 GHz respectively. Accurate, versatile, affordable, rugged and a true handheld at under 5 pounds (including battery), Site Master is ideally suited for field technicians involved with installing, provisioning, maintaining, and troubleshooting wireless base station cable and antenna systems for such applications VHF, broadcasting, paging, cellular, PCS/GSM, 3G, ISM, WLAN and WLL. The Site Master S332D cable and antenna analyzer features spectrum analysis capability and optional power meter capability in addition to Return Loss/ SWR (also known as VSWR, for Voltage Standing Wave Ratio) and Cable Loss measurements, as well as Distance to Fault (DTF) analysis using Frequency Domain Reflectometer (FDR). These capabilities give field engineers and technicians the ability to easily identify and solve RF system problems such as coverage, interference, and other path related signal problems. S332D spectrum analysis ranges from 100 kHz to 3 GHz with smart button measurements – Field Strength, Occupied Bandwidth (OBW), Channel Power, Adjacent Channel Power Ratio (ACPR), Interference Analysis and Carrier-to-Interference ratio (C/I). Other features include <3.5 msec Sweep Speed Per Data Point to enable identification of intermittent problems in real time; broadband calibration from 25 MHz to 4 GHz for troubleshooting cable and antenna systems without multiple calibrations and calibration setups; and built-In Worldwide Signal Standards eliminating the need to manually input start and stop frequency values. The High Accuracy Power Meter (Option 19) provides true RMS measurements from 50 MHz to 6 GHz. This is the ideal solution for accurate measurements of both CW and modulated signals.

 

Features

  • Handheld, battery-operated design
  • Weighs less than 5 lbs. (including battery)
  • Built-in worldwide signal standards and frequency channels
  • Superior Immunity to RF interference
  • Locate long range problems with 517 data points
  • Pop-up cable menus containing over 95 cable types
  • Multilingual User Interface; English, French, Chinese, Japanese, Spanish, German
  • Intuitive and easy to use with On-Screen Test Set-Ups
  • Saves 15 to 35 Measurement Test Set-Ups (10 VNA, 5 SPA/Opt 21, 5 each Option 19/25/27/29)
  • Saves up to 300 Measurement Traces
  • Alphanumeric Labeling of Saved Measurements
  • Automatic Time and Date Stamp of Saved Data
  • Rechargeable, Field Replaceable Battery
  • Distance Measurement in Feet and Meters
  • Information Screens Provide Cable Attenuation and Velocity of Propagation values
  • 6 Markers, Limit Lines, and Segmented Limit Lines
  • RS232 Interface
  • Built in pre-amplifier (standard)
  • Noise Level < – 135 dBm typical
  • Resolution Bandwidth: 100 Hz to 1 MHz in 1-3 sequence
  • Video Bandwidth: 3 Hz to 1 MHz in 1-3 sequence
  • Dedicated smart measurements
  • + 43 dBm maximum safe input level

Anritsu S332D Applications

  • Cable and antenna testing for VHF, broadcasting, paging, cellular, PCS/GSM, 3G, ISM, WLAN and WLL applications.
  • Return Loss measurement
  • Distance-To-Fault measurement
  • Cable loss measurement Interference Detection and Analysis
  • Rapid sweep speed to detect interference
  • AM/FM/SSB detection
  • Interfering signal type determination helps quickly determine source of cellular interference Base Station Transmitter Performance Testing
  • Channel power
  • Adjacent Channel Power Ratio
  • Occupied Bandwidth
Print
Email

Specifications

Frequency Range

25 to 4000 MHz

Frequency Accuracy

<±50 ppm @ +25°C

Frequency Resolution

1 kHz (CW On)

100 kHz (CW 0ff)

Output Power

0 dBm (typical)

Number of Data Points

130 or 259 or 517

VSWR:

Range

1.00 to 65.00

 

Resolution

0.01

Return Loss

Range

0.00 to 60.00 dB

 

Resolution

0.01 db

Test Port, Type N 50 Ohms ***Immunity to Interfering signals up to the level of –5 dBm, reflection

Maximum Input (Damage Level)

Test Port, Type N: +22 dBm

RF Detector:

+20 dBm

Measurement Accuracy

Measurement accuracy depends on calibration components

> 42 dB corrected directivity after calibration

Cable Loss

Range

0.00 to 30.00 dB

Resolution

0.01 dB

Frequency Reference

Aging

±1 ppm/yr

Accuracy

±2 ppm

Internal Memory:

Trace Memory

300 maximum

Instrument configuration

10 setup locations

RS-232: 9 pin D-sub, three wire serial

Electromagnetic Compatibility Meets European community requirements for CE marking

Power Supply

External DC Input:  +11 to +15 Vdc, 1250 mA max

Temperature

Storage

-51° C to 71° C

Operation

-10 C to 55° C

Weight

4 lbs (1.82 kg)

Dimensions

10 x 7 x 2.4 in (25.4 x 17.8 x 6.1 cm)

Stay informed

To keep you in the loop of our latest news, register now for our email newsletter.