Associated Research 620L - Fully Automated Line Leakage Tester
Description
The 620L is a standalone Line Leakage Tester with an enhanced graphic LCD which automates leakage testing in production and laboratory environments. The 620L is configured for up to 40 Amps of current draw for DUT input power. It is designed to test to most safety agency standards for Line Leakage testing. The 620L comes standard with USB and RS-232 interfaces. Ethernet, GPIB, and RS-485 interfaces are also available.
Features
- Test operators can configure the 620L to perform all eight required Line Leakage tests
- Leakage current readings can be monitored using both PEAK and RMS measurements
- Most common measuring devices are already incorporated into the instrument’s intuitive menu system
- 50 Memories with 30 steps per memory can be stored and recalled in any alphanumeric combination
- Compact 3U Rack Mount Design
- Optional Functional Run Testing for additional Measurements
- Interconnection to APT Brand AC Power Source
- Interconnection to SC6540 Modular Scanner provides automated control of multiple test points
- Graphic LCD and intuitive menu system to simplify the entire testing process
- Patented CAL-ALERT® alerts the operator that the 620L is due for re-calibration
- Handles up to 40 Amp maximum continuous DUT Current
- Optional cold resistance measurement capability
- USB/RS-232, GPIB, Ethernet, or RS-485 automation interfaces available
- Easily Interconnect to any automated Associated Research Hipot Tester
- Autoware Testing Software available for complete Automation Control
Specifications
Input Specifications |
Voltage |
115/230 VAC ± 10%, user selection |
||||
Frequency |
50/60 Hz ± 5% |
|||||
Fuse |
2 A Slow Blow 250 VAC |
|||||
Line Conditions |
Reverse Power |
Switch Reverse polarity switch for normal condition |
||||
Neutral Switch |
Neutral switch on/off selection for single fault |
|||||
Ground Switch |
Ground switch on/off selection for class I single fault |
|||||
Probe Settings |
Surface to Surface |
(PH – PL) |
||||
Surface to Line |
(PH – L) |
|||||
Ground to Line |
(G – L) |
|||||
Leakage Limit Settings |
Touch Current High/Low Limit (RMS) |
Range |
0.0 μA – 999.9 μA / 1000 μA – 9999 μA /10.00 mA – 20.00 mA |
|||
Resolution |
0.1 μA/1 μA/0.01 mA |
|||||
Touch Current High/Low Limit (PEAK) |
Range |
0.0 μA – 999.9 μA/ 1000 μA – 9999 μA / 10.00 mA – 30.00 mA |
||||
Resolution |
0.1 μA/1 μA/0.01 mA |
|||||
Display |
Touch Current Display (RMS) |
Range |
0.0 μA – 550 μA, frequency DC, 15 Hz – 1 MHz |
|||
Resolution |
0.1 μA |
|||||
Accuracy |
DC: 15 Hz < f < 100 kHz: ±(2% of reading + 3 counts) 100 kHz < f < 1 MHz: ±5% of reading, (10.0 μA – 999.9 μA) |
|||||
Range |
400 μA – 8500 μA, frequency DC, 15 Hz – 1 MHz |
|||||
Resolution |
1 μA |
|||||
Accuracy |
Accuracy: DC: 15 Hz < f < 100 kHz: ±(2% of reading + 3 counts) 100 kHz < f < 1 MHz: ±5% of reading, (10 μA – 8500 μA) |
|||||
Range |
8.00 mA – 20.00 mA, frequency DC, 15 Hz – 1 MHz |
|||||
Resolution |
0.01 mA |
|||||
Accuracy |
DC: 15 Hz < f < 100 MHz: ±5% of reading, (0.01 mA -20.00 mA) |
|||||
Touch Current Display (PEAK) |
Range |
0.0 μA – 550 μA, frequency DC – 1 MHz |
||||
Resolution |
0.1 μA |
|||||
Accuracy |
DC : ±(2% of reading +3 counts) 15 Hz < f < 1 MHZ : ±10% of reading +2 μA |
|||||
Range |
400 μA – 8500 μA, frequency DC – 1 MHz |
|||||
Resolution |
1 μA |
|||||
Accuracy |
DC: ±(2% of reading +3 counts) 15 Hz < f < 1 MHz ±10% of reading +2 μA |
|||||
Range |
8.00 mA – 30.00 mA, frequency DC – 100 kHz |
|||||
Resolution |
0.01 mA |
|||||
Accuracy |
DC: ±(2% of reading +3 counts) 15 Hz < f < 100 kHz ±10% of reading +2 counts |
|||||
General Specifications |
Dimension (W x H x D) |
16.93 x 5.24 x 11.81 in (430 x 133 x 300 mm) |
||||
Weight |
26.45 lbs (12 kg) |
|||||
Display |
320 X 240 graphic LCD |
|||||
Mechanical |
Bench or rack mount with tilt up feet |
|||||
Memory |
50 Memories, 30 steps per each memory File locations can link 900 steps max |
|||||
Interface |