Bruel & Kjaer Type 2250-H Hand-Held Vibration Analyzer
Description
Type 2250-H is suitable for a wide variety of vibration measurement applications using one of a range of accelerometers – dependent on the level, position and environment in which the measurements are to be made. Type 2250 H includes the analysis tools and measurement units for a wide range of troubleshooting applications. Measuring the vibration level of a pump, fan or motor is a simple and effective technique for determining the correct operation of the machine or component. Higher-than-expected vibration magnitudes may indicate possible problems with shaft misalignment, loosening of components or general wear on bearings and gears.
Features
- Large, high-resolution, touch-sensitive color screen
- Traffic light status indicator
- Plug-in rechargeable Li-ion battery
- Data storage on plug-in SD and CF memory cards
- 120 dB dynamic range
- Wireless LAN connectivity
- Vibration signal recording for post-processing analysis
- Robust and environmentally protected (IP 44)
Specifications
|
SUPPLIED MICROPHONE |
|
Type 4189: Prepolarized Free-field ½″ Microphone |
|
Nominal Open-circuit Sensitivity: 50 mV/Pa (corresponding to −26 dB re 1 V/Pa) ± 1.5 dB |
|
Capacitance: 14 pF (at 250 Hz) |
|
MICROPHONE PREAMPLIFIER ZC0032 |
|
Nominal Preamplifier Attenuation: 0.25 dB |
|
Connector: 10-pin LEMO |
|
Extension Cables: Up to 100 m in length between the microphone preamplifier and Type 2250, without degradation of the specifications |
|
Accessory Detection: Windscreen UA 1650 can be automatically detected when fitted over ZC0032 |
|
MICROPHONE POLARIZATION VOLTAGE |
|
Selectable between 0 V and 200V |
|
KEYBOARD |
|
Pushbuttons: 11 keys with backlight, optimized for measurement control and screen navigation |
|
ON-OFF BUTTON |
|
Function: Press 1 s to turn on; press 1 s to enter standby; press for more than 5 s to switch off |
|
STATUS INDICATORS |
|
LEDs: Red, amber and green |
|
DISPLAY |
|
Type: Transflective back-lit color touch screen 240 × 320 dot matrix |
|
Color Schemes: Five different optimized for different usage scenarios (day, night, etc.) |
|
Backlight: Adjustable level and on-time |
|
USER INTERFACE |
|
Measurement Control: Using pushbuttons on keyboard Setup and Display of Results: Using stylus on touch screen or pushbuttons on keyboard |
|
Lock: Keyboard and touch screen can be locked and unlocked |
|
USB INTERFACE |
|
USB 1.1 OTG Mini B socket |
|
INPUT SOCKET |
|
Connector: Triaxial LEMO |
|
Input Impedance: ≥1MΩ |
|
Signal Source Impedance: < 200Ω |
|
Direct Input: Max. input voltage: ± 14.14Vpeak |
|
CCLD Input: Max. input voltage: ± 7.07 Vpeak |
|
CCLD Current/voltage: 4mA/25V |
|
TRIGGER SOCKET |
|
Connector: Triaxial LEMO |
|
Max. Input Voltage: ± 20Vpeak |
|
Input Impedance: > 1MΩ |
|
OUTPUT SOCKET |
|
Connector: Triaxial LEMO |
|
Max. Peak Output Level: ± 4.46V |
|
Output Impedance: 50Ω |
|
HEADPHONE SOCKET |
|
Connector: 3.5 mm Minijack stereo socket |
|
Max. Peak Output Level: ± 1.4V |
|
Output Impedance: 2.2Ω in each channel |
|
MICROPHONE FOR COMMENTARY |
|
Microphone, which utilizes Automatic Gain Control (AGC), is incorporated in underside of instrument. Used to create voice annotations for attaching to measurements |
|
EXTERNAL DC POWER SUPPLY REQUIREMENTS |
|
Used to charge the battery pack in the instrument |
|
Voltage: 8 . 24 VDC, ripple voltage < 20mV |
|
Current Requirement: min. 1.5 A |
|
Power Consumption: < 2.5W, without battery charging, < 10W when charging |
|
Cable Connector: LEMO Type FFA.00, positive at centre pin |
|
BATTERY PACK |
|
Type: Li-Ion rechargeable |
|
Typical Operating Time: > 8 hours |
|
STORAGE SYSTEM |
|
Internal Flash-RAM (non-volatile): 20 Mbyte for user setups and measurement data |
|
External Secure Digital Memory Card (SD-card): For store/recall of measurement data |
|
External Compact Flash Memory Card (CF-card): For store/recall of measurement data |
|
CLOCK |
|
Back-up battery powered clock. Drift <0.5s per 24 hour period |
|
WARM-UP TIME |
|
From Power Off: < 2 minutes |
|
From Standby: < 10 seconds for prepolarized microphones |
|
TEMPERATURE |
|
IEC 60068.2.1 & IEC 60068.2.2: Environmental Testing. Cold and Dry Heat. |
|
Operating Temperature: .10 to + 50°C (14 to 122°F), < 0.1 dB |
|
Storage Temperature: .25 to +70°C (.13 to 158°F) |
|
HUMIDITY |
|
IEC 60068.2.78: Damp Heat: 90% RH (non-condensing at 40°C (104°F)) |
|
Effect of Humidity: < 0.1 dB for 0% < RH < 90% (at 40°C (104°F) and 1 kHz) |
|
MECHANICAL |
|
Environmental Protection: IP44 |
|
Non-operating: |
|
IEC 60068.2.6: Vibration: 0.3 mm, 20 m/s2, 10 . 500Hz |
|
IEC 60068.2.27: Shock: 1000 m/s2 |
|
IEC 60068.2.29: Bump: 4000 bumps at 400 m/s2 |
|
WEIGHT AND DIMENSIONS |
|
650 g (23 oz.) including rechargeable battery |
|
300 × 93 × 50mm (11.8 × 3.7 × 1.9″) including preamplifier and microphone |
|
LANGUAGE |
|
User Interface in Catalan, Croatian, Czech, Danish, English, Flemish, French, German, Hungarian, Japanese, Italian, Polish, Romanian, Serbian, Slovenian, Spanish and Swedish |
|
HELP |
|
Concise context-sensitive help in Catalan, English, French, German, Italian, Polish, Romanian, Serbian, Slovenian and Spanish |