QuadTech Hybrid 2000 AC/DC Hipot, IR, DCR 5 kVAC - 6 kVDC
Description
The Hybrid 2000 Wound Component Scanner is a production safety tester designed for test requirements of three-phase motors, double-wound transformers and heater related wound components. Quality verification test items for wound components consist of AC/DC Hipot (dielectric withstanding), Insulation Resistance (IR), and Impulse Winding tests. We have integrated these tests into one unit.
Features
- Open/Short check (OSC) function to ensure of DUT connection
- SUB-STEP function
- GFI Human protection – shutdown if current imbalance >0.5mA
- DWX Series – Impulse Winding Tester compatible
- Flashover detection
- Key Lock function
- RS232 Interface (standard)
- GPIB & Handler (optional)
- Friendly interface
- CE Mark
Specifications
Hybrid 2000 AC/DC Hipot, IR, DCR |
||
AC Output Voltage |
Range |
50V to 5000V AC |
Resolution |
1Volts/step |
|
Frequency |
50/60 Hz selectable |
|
Waveform |
Sinusoidal |
|
Voltage Display |
Accuracy |
±(1% of reading + 5V) |
Resolution |
1 Volt |
|
DC Current Display |
Range |
0.01mA to 30mA AC |
Resolution |
1μA |
|
Accuracy |
±(1% or reading + 5cnt) Total |
|
High/Low Limit Test |
0.001mA to 30mA AC Low limit can be turned OFF |
|
Insulation Resistance |
Range |
1MΩ – 50GΩ |
Accuracy |
1M – 1G: ±(3% + 10 cts), >500V |
|
|
1G – 10G: ±(7% + 10 cts), >500V |
|
|
10G – 50G: ± (10% + 10 cts), >500V |
|
IR Output Voltage: |
Range |
50V to 5000V DC |
|
Accuracy |
±(1.5% of setting +5V) |
High/Low Limit Test |
1MΩ – 50GΩ |
|
Low limit can be turned OFF |
||
DC Resistance |
Range |
10mΩ – 100kΩ |
DCR Balance |
0.0001Ω – 999Ω, OFF |