EM TEST ACS 500N6 Single Phase AC Voltage Source 6 kVA

Description

EM Test ACS 500N6 is a single phase AC source specifically designed for harmonics and flicker testing. It meets the corresponding specifications as per EN/IEC 61000-3-2 and EN/IEC 61000-3-3 as well as JIS C61000-3-2.

As per standard requirements a pure voltage is needed for measurements of harmonics and flicker. The output voltage of the ACS 500N6 is guaranteed to be sinusoidal with a very low distortion (THD) of less than 0.1% regardless of the load.

The ACS 500N6 is electronically protected against Overtemperature, Overcurrent, Overvoltage and Undervoltage.

Features

  • Frequency (fund.) 10Hz .. 80Hz
  • Output power max. 6,000VA
  • Output voltage max. 300V
  • Inrush current capability
  • Remote control via RS232
Download Data Sheet
Print
Email

Specifications

OUTPUT
Voltage range 0V to 300V
Voltage resolution 0.025% (12 bit)
Output frequency 10Hz to 80Hz
Frequency resolution 0.02Hz
Frequency accuracy, stability 100ppm
Output power 6,000VA
Output Safety laboratory connector
Output current, 50Hz Continuous: 24A maximum

RMS (max. 3s): 47A

Repetitive peak: 200A

Output current, 60Hz Continuous: 24.9A maximum

RMS (max. 3s): 47.9A

Repetitive peak: 200A

INPUT
Supply voltage 3 x 400V (3P,N,PE)
Input current 32A (16A/27A neutral)
Line frequency 45Hz – 65Hz
Fuse 3x 16A internal
Connector CEE type 32A
CONTROL
Interface RS 232, 9-pin SubD
DIMENSIONS
Housing 19″, 6HU
Weight 40kg
Size 287mm x 449mm x 500mm
Stay informed