EM TEST AIF 503N-series 3-Phase Flicker Impedances 16A – 75A
Description
EM TEST AIF 503N-series 3-Phase Flicker Impedances 16A – 75A unifies the required impedance values for two applications (except for the AIF 503N16 model). It offers a flicker impedance with the values in accordance with IEC 60725 for flicker measurements as per IEC/EN 61000-3-3. It also includes a second impedance used for flicker measurements on DUTs with a rated current above 16A which are intended to be connected to a public low-voltage distribution system with a service current of more than 100A.
The inductors used are non-saturable air coils and are manually matched to the specified value.
The AIF 503N-series is used along with the DPA503N for single phase as well as for three phase measurements.
MODEL OVERVIEW
AIF 503N16 Three-Phase Flicker Impedance 16A
AIF 503N32.1 Three-Phase Flicker Impedance 32A
AIF 503N63.1 Three-Phase Flicker Impedance 63A
AIF 503N75.1 Three-Phase Flicker Impedance 75A
Features
- Impedances values as per IEC/EN 61000-3-3 & IEC/EN 61000-3-11
- Non-saturable air coil inductors
- Line voltage 3 x 400V
- Line current up to 75A per phase (depending on model)
- Automatic overcurrent protection
- Status indication by LED
- Remote-controlled by DPA 503N
Specifications
COMMON DATA (ALL MODELS) | |
GENERAL | |
Configuration | 3-phase (3L/N/PE) |
Line voltage | 3 x 400V ± 10% |
Line frequency | 47Hz – 63Hz |
Bypass for harmonics | Remote controlled switch |
Interface | 15-pin SubD (DPA 503N control) |
Indicators | LED’s indicating the active status (Bypass or impedance in use). |
IMPEDANCE | |
Zref at 50Hz | RA = 0.24ohm, XA = 0.15ohm
RN = 0.16ohm, XN = 0.10ohm |
Ztest at 50Hz | RA = 0.15ohm, XA = 0.15ohm
RN = 0.10ohm, XN = 0.10ohm (not available in AIF 503N16) |
Accuracy | Better than 3%
Each inductor is a non-saturable air coil, matched manually to the specified value. |
AMBIENT CONDITIONS | |
Temperature | 0°C to 35°C |
Relative humidity | 80%, non-condensing |
Atmospheric pressure | 86 kPa (860 mbar) to 106 kPa (1,060 mbar) |
Cooling | fans, active cooling control |