EM TEST RCB 200N1 Transient Immunity Test Generator
Description
RCB 200N1 Transient Immunity Test Generator meeting Ford FMC1278 and EMC-CS-2009.1
EM TEST RCB 200N1 Transient Immunity Test Generator is built to generate pulses such as CI 220 A1, A2-1, A2-2, C-1 and C-2, CI 260 F and RI 130 A2-1 and A2-2 for which the RCB 200N1 is built into a metal housing. For RI 130 testing the RCB 200N1 offers BNC-type coaxial outputs to connect the specified test fixture needed for this test.
Features
- Approved by Ford acc. to FMC1278 and EMC-CS-2009.1 requirements
- Pulses CI 220: A1, A2-1, A2-2, C-1 and C-2
- Pulse CI 260: F
- Pulses RI 130: A2-1, A2-2 using a coupling test fixture
- Fixed repetition times and pseudo-random timing
- Automatic control of mode and switch settings by micro-controller
- Easy control by manual operation
- EOL status control of the relay as per FORD requirement
- Internal/external trigger mode operation
Download Data Sheet
Print
Email
Specifications
| TECHNICAL DATA | |
| RCB 200N1 | Circuitry as per Annex F, Figure F-1 with the following components: |
| Resistor R1 | 51 ohms, 25W |
| Resistor R2 | 220 ohms ± 5%, 2W |
| Resistor R3 | 33 ohms ± 5%, 10W |
| Resistor R4 | 6 ohms ± 5%, 50W |
| Capacitor C1 | 100 nF capacitor, 400V |
| Inductor L1 | 5 uH inductor (Osborne transformer part no. 8745) |
| Inductor L2 | 100 mH inductor (Osborne transformer part no. 32416) |
| Diode D1 | Zener diode, 39V, 5W |
| Transistor Q1 | NPN transistor |
| SW0 – SW4 | Single Throw Switch |
| RLY1 | 12-volt AC relay, NC contact used Potter & Brumfield KUP-14A15-12) |
| OPERATION | |
| TEST ON | Activates the DUT supply for CI 220 and CI 260 test pulses; no function for RI 130 as this is performed without DC voltage |
| ID | To select the test |
| Pulse | To select the pulses of the preset test |
| Mode | To select the mode of generating the pulses |
| Start/Stop | To start/stop the test |
| INPUTS/OUTPUTS | |
| Input | 4mm safety lab connectors for DC supply (e.g. battery or VDS 200x) |
| Trigger | External trigger input (BNC) |
| Outputs | 4mm safety lab connectors for CI 220 and CI 260 testing; 50-ohm BNC connectors for RI 130 testing |
| LEDs | Indicates:
– TEST ON status – Test (ID) – Pulse – Mode – Sequence or endless – Switching status (SW0 – SW4) – Relay EOL (end of life reached) |
| GENERAL DATA | |
| Dimensions
Weight |
330mm x 230mm x 115mm
Approx. 7.5kg |
| Fuse | 20A for DUT supply current |
| Protection | Electronically protected against overvoltage and reversed polarity |