EM TEST UCS 500N5 Transient Generator

Description

UCS 500N5 Multifunctional Test Generator for Transients (EFT/Burst, Surge & Power Fail) up to 5.5 kV as per IEC/EN 61000-4-4, 61000-4-5 and 61000-4-11

EM TEST UCS 500N5 Transient Generator is the most versatile tester to cover transient and power fail requirements according to international standards (basic and generic standards) and product/product family standards. The UCS 500N5 is the most economical solution for tests during development as well as for full-compliant immunity tests and CE Marking for single phase DUT with the ability to be extended for testing three-phase DUTs by means of an automatically controlled external coupling network up to 100A.

 

MODEL OVERVIEW

UCS 500N5E Electrical Fast Transient/Burst Generator up to 5.5 kV as per IEC/EN 61000-4-4
UCS 500N5P Power Fail Simulator for Dips & Short Interruptions as per IEC/EN 61000-4-11 and 61000-4-29
UCS 500N5V Combination Wave (Surge only) Generator up to 5kV as per IEC/EN 61000-4-5

Features

  • Burst module (IEC/EN 61000-4-4) up to 5.5kV
  • Surge module (IEC/EN 61000-4-5) up to 5kV
  • PowerFail module (IEC/EN 61000-4-11)
  • Magnetic field tests with optional accessories
  • Built-in single phase coupler 300V/16A
  • Manual operation from front panel
  • USB and GPIB-Bus for remote control
  • Monitor outputs for peak voltage and peak current
  • Fail inputs
  • External variac control 0-10Vdc
  • Control of external three-phase couplers
Download Data Sheet
Print
Email

Specifications

TECHNICAL DETAILS
Burst Module EFT/N5 as per EN/IEC 61000-4-4 and EN 61000-6-1, -6-2
Test voltage 200V – 5,500V ± 10%;

100V – 2,750V ± 10% into 50ohm

Pulse shape 5/50ns into 50ohm and 1,000ohm
Rise time tr 5ns ± 30% into 50ohm;

5ns ± 30% into 1,000ohm

Pulse width td 50ns ± 30% into 50ohm;

50ns -15/+100ns into 1,000ohm

Source impedance 50ohm
Polarity Positive/negative
COMBINATION WAVE/SURGE
Surge Module VCS/N5 as per EN/IEC 61000-4-5 and EN 61000-6-1, -6-2
Voltage (o.c.) 160V – 5,000V ± 10%
Pulse front time 1.2us ± 30%
Pulse time to half value 50us ± 20%
Current (s.c.) Max. 2,500A ± 10%
Pulse front time 8us ± 20%
Pulse time to half value 20us ± 20%
Polarity Positive/negative/alternating
Event counter 1 – 30,000 or endless, selectable
POWER FAIL, DIPS & INTERRUPTS, VOLTAGE VARIATIONS
Power Fail Module PFS/N5 as per EN/IEC 61000-4-11, IEC/EN 61000-4-29 and EN 61000-6-1, -6-2
Channel PF1/PF2 AC voltage: max. 300V

AC current: max. 16A

DC voltage: max. 300V

DC current: max. 16A

Frequency 16Hz – 500Hz and DC
Switching time < 5us into a 100ohm resistive load
Inrush current > 500A
Protection Both channels are protected against short-circuit conditions.
GENERAL DATA
Interfaces
Serial interface USB
Parallel interface IEEE 488, addresses 1 – 30
Analog output 0 – 10VDC to control an external transformer
CN interface 15pin SubD connector to control an external coupling network
Fail inputs DUT monitoring via Fail1 and Fail2 input (one each)
Dimensions
Dimensions 19″, 3HU
Weight approx. 25kg