EM TEST UCS 500N7.2 Transient Simulator
Description
UCS 500N7.2 – Enhanced Ultra Compact Tester as per IEC 61000-4-x and ANSI/IEEE C62.41 for Renewable Energy
EM TEST UCS 500N7.2 Transient Simulator covers transient and power fail requirements according to international standards (basic and generic standards) and product/ product family standards with voltage capability of up to 7.0 kV. Apart from the IEC/EN 61000-4-5 standard for surge testing it also complies to ANSI/IEEE C62.41 for surge testing.
The UCS 500N7.2 conducts fully compliant tests at levels that go far beyond common test requirements. It has a built-in CDN for single phase DUTs which can be extended for testing three-phase DUTs by means of an automatically controlled external coupling network up to 100A. Surge can be tested up to 7kV according to IEC 61000-4-5 or ANSI/IEEE C62.41.
The UCS 500N7.2 is an economical solution for full-compliant immunity tests and CE Marking.
MODEL OVERVIEW
UCS 500N7 Enhanced Ultra Compact Tester as per IEC 61000-4-x and ANSI/IEEE C62.41
Features
- Ultra-Compact Simulator up to 7.0kV
- Burst module (IEC/EN 61000-4-4)
- Surge module (IEC/EN 61000-4-5)
- PowerFail module (IEC/EN 61000-4-11)
- Telecom Surge Module (IEC/EN 61000-4-5) optional
- Magnetic field tests with optional accessories
- Built-in single-phase coupler 400V/16A
- USB and GPIB-Bus for remote control
- Manual operation from front panel
- Monitor outputs for peak voltage and peak current
- Fail inputs
- External variac control 0-10Vdc
- Control of external three-phase couplers
Specifications
| ELECTRICAL FAST TRANSIENTS | |
| Burst Module EFT/N7 | As per EN/IEC 61000-4-4 and EN 61000-6-1, -6-2 |
| Test voltage | 200V – 5,500V ± 10%;
100V – 2,750V ± 10% into 50ohm |
| Pulse shape | 5/50ns into 50ohm and 1,000ohm |
| Rise time tr | 5ns ± 30% into 50ohm;
5ns ± 30% into 1,000ohm |
| Pulse width td | 50ns ± 30% into 50ohm;
50ns -15/+100ns into 1,000ohm |
| Source impedance | 50ohm |
| Polarity | Positive/negative |
| COMBINATION WAVE / SURGE | |
| Surge Module VCS/N7 | As per EN/IEC 61000-4-5 and EN 61000-6-1, -6-2 |
| Voltage (o.c.) | 250V – 7,000V ± 10% |
| Pulse front time | 1.2us ± 30% |
| Pulse time to half value | 50us ± 20% |
| Current (s.c.) | Max. 3,500A ± 10% |
| Pulse front time | 8us ± 20% |
| Pulse time to half value | 20us ± 20% |
| Polarity | Positive/negative/alternating |
| Event counter | 1 – 30,000 or endless, selectable |
| POWER FAIL, DIPS & INTERRUPTIONS, VOLTAGE VARIATIONS | |
| Power Fail Module PFS/N7 | As per EN/IEC 61000-4-11, and EN 61000-6-1, -6-2 |
| Channel PF1/PF2 | AC voltage: max. 400V
AC current: max. 16A |
| Frequency | 16Hz – 500Hz |
| Switching time | < 5us into a 100ohm resistive load |
| Inrush current | > 500A |
| Protection | Both channels are protected against short-circuit conditions. |
| GENERAL DATA | |
| INTEGRATED COUPLING/DECOUPLING NETWORK | |
| Dut Supply | Max. 400V/16A (AC): 50/60 Hz
Max. 300V/16A (DC) |
| INTERFACES | |
| Serial interface | USB |
| Parallel interface | IEEE 488, addresses 1 – 30 |
| Analog output | 0 – 10VDC to control an external transformer |
| CN interface | 15pin SubD connector to control an external coupling network |
| Fail inputs | DUT monitoring via Fail1 and Fail2 input (one each) |
| MAINS | |
| Supply voltage | 115 / 230VAC + 10%/-15% |
| Power consumption | Approx. 75W |
| Frequency | 50/60 Hz |
| Fuses | 2xT 2A (230V) or 2xT 4A (115V) |
| DIMENSIONS | |
| Dimensions | 19″, 6HU, |
| Weight | approx. 29kg |
- ANSI/IEEE Standards
- ANSI/IEEE C62.41
- Basic Standards
- IEC 61000-4-11
- IEC 61000-4-29
- IEC 61000-4-4
- IEC 61000-4-5
- IEC 61000-4-8
- IEC 61000-4-9
- Telecom Standards
- ITU-T K.20
- ITU-T K.21
- ITU-T K.45
- Product Standards
- EN 61543
- IEC 61008-1
- IEC 61009-1
- IEC 61326-1
- IEC 61543
- IEC 61850-3
- UL Standards
- UL 1741
- Generic Standards
- EN 61000-6-1
- EN 61000-6-2
- Automotive OEM Specifications
- UNECE R10 (Automotive)
- International Standards
- IEEE 1547