ETS-Lindgren HI-6053 Electric Field Probe 10 MHz – 40 GHz

Description

The battery-powered HI-6053 Electric Field Probe provides broadband frequency coverage and wide dynamic range that satisfies the demands of most test requirements. To take advantage of this capability, the HI-6053 was designed to be single range reading so data can be read continuously over the entire dynamic range. Data values for each axis (X, Y, and Z) can be read individually or summed.

The HI-6053 Electric Field Probe is a fully intelligent sensor enabling fast and accurate EMF measurements with industry-leading performance specifications. Optical coupling to a variety of readout options makes this probe ideally suited for a wide range of field monitoring applications.

The HI-6100 Field Monitor provides manual functions and programmed control via IEEE-488 and RS-232 Serial Data Interfaces. Readings from up to four probes can be displayed simultaneously.

For direct connection to a PC either the HI-4413P Serial or the HI-4413USB interfaces are available.

This field probe incorporates user replaceable/rechargeable batteries.

Features

  • Frequency Range: 10 MHz – 40 GHz
  • Provides Individual and Summed Axis Values
  • Dynamic Range: 2.0 – 800 V/m (single range)
  • User replaceable/rechargeable batteries
  • A2LA Accredited Calibration Report
  • Suitable for MIL Standard Specs:
    — MIL-STD 461F Radiated Susceptibility (RS)
  • Suitable for Automotive Specs:
    — SAE J1113/27
    — GM9120P
    — GM93100GS
    — GM9114P
  • Operates with most 3rd Party Immunity software
Download Data Sheet
Print
Email

Specifications

Electrical Specifications
Detection Isotropic (X, Y, and Z Axis Readings)
Frequency Response with Correction 10 MHz to 18 GHz ± 0.9 dB
18 to 40 GHz ± 1.1 dB
Frequency Range 10 MHz to 40 GHz
Frequency 10 to 100 MHz + 3.0, -4.0 dB
100 MHz to 1 GHz +3.0, -0.50 dB
1 to 18 GHz +4.0, -2.0 dB
18 to 40 GHz +3.5, -4.5 dB
Dynamic Range 2 to 800 V/m (Single Range)
Resolution 0.01 V/m
Isotropic Deviation ±1.0 dB < 18 GHz
Linearity ±0.5 dB @ 1 GHz
Sample Rate (Typical) HI-4413P Serial Modem – 44 samples/second maximum
HI-6100 Field Monitor – 70 samples/second maximum
Overload Withstand >1500 V/m CW
Physical Specifications
Physical Interface FSMA Connectors
Duplex Optical Fiber (200 micron multimode)
Battery Rechargeable NiMH AAA
Operating Temperature Range 10° to 40° C, 5-95% Relative Humidity Non-condensing
Battery Life >30 Hours Continuous
Battery Charger 100-240 VAC Universal Input
3-Hour Charge from Full Depletion
Weight 0.36 kg (12.64 oz)
Mounting 1/4-20 UNC Internal Thread
Dim. L. 43.8 cm   (17.24 inches)
Dim. W. 5.7 cm   (2.24 inches)
Stay informed