ETS-Lindgren HI-6122 Electric Field Probe 10 kHz – 1 GHz

Description

The laser powered HI-6122 EMF Field Probe provides broadband EMF frequency coverage and wide dynamic range that satisfies the demands of most test requirements. To take advantage of this capability, the HI-6122 was designed to be single range reading so data can be read continuously over the entire dynamic range. Data values for each axis (X, Y, and Z) can be read individually or summed.

Fiber optic signal and power lines link the Model HI-6122 probe to either the model HI-6100 Field Monitor or as a direct connect to a PC with the HI-6113 Laser Data Interface and ProbeView Laser Software.

The HI-6100 provides manual functions and programmed control via IEEE-488 and RS-232 Serial Data Interfaces. Readings from up to four probes can be displayed simultaneously.

The HI-6113 Laser Data Interface provides the laser power and communications for the HI-6122 Electric Field Probe. A USB connection to the PC allows for quick and easy data collection, using ProbeView Laser software.

Features

  • Laser Powered – permits extended testing
  • Frequency Range: 10 kHz – 1 GHz
  • Dynamic Range 2.0 to 800 V/m (Single Range)
  • Provides Individual and Summed Axis Values
  • A2LA Accredited Calibration Report
  • Suitable for MIL Standard Specs:
    — MIL-STD 431E Radiated Susceptibility (RS)
  • Suitable for Automotive Specs:
    — SAE J1113/27
    — GM9120P
    — GM93100GS
    — GM9114P
  • Suitable for Commercial Specs:
    — EN/IEC61000-4-3 Radiated Immunity
  • Operates with most 3rd Party Immunity software
Download Data Sheet
Print
Email

Specifications

Electrical Specifications
Detection Isotropic (X, Y and Z Axis Readings)
Frequency Range 10 kHz – 1 GHz
Frequency Response with Correction 10 kHz to 1 GHz ± 0.9 dB
Frequency Response 10 kHz to 30 kHz – 2.5/+0.5 dB
30 kHz – 1 GHz +/- 1.0 dB
Dynamic Range 2.0 to 800 V/m (Single Range)
Sensor Protection Caps Height 43 mm
Resolution 0.01 V/m
Isotropic Deviation ± 0.5 dB @ 400 MHz dB
Linearity ± 0.5 dB @ 27 MHz
Sample Rate (Typical) > 70 Samples second max.
Overload Withstand >1500 V/m CW
Physical Specifications
Physical Interface Duplex Optical Fiber (62.5 micron multimode)
FC Connectors for Laser cable, integral 1m optical cable
ST Connector for Transmitter cable, Integral 1m optical cable
Power Laser-powered
Operating Temperature Range 10 to 40° C
Dimensions 32 mm x 32 mm x 32 mm Housing
(1.25 in. x 1.25 in. x 1.25 in.)
36 mm (1.42 in) Sensor Protection Caps
Weight 80 g   ( 2.82 ounces )