Fluke 5215A Precision Power Amplifier

Description

The Fluke 5215A Precision Power Amplifier is an ac coupled programmable inverting amplifier with a fixed gain of 100. Output level is specified to 1100V at up to 200 mA, with a typical upper limit of 1200V before automatically tripping into standby mode.

The 5215A includes automatic-overload sensing and recovery. Upon sensing an excessive slew rate or frequency of the input signal, or a momentary output overload, output is returned to zero within 2 microseconds and held there for 6 milliseconds, or until the fault is corrected. When steady overload, shorted output, or excessive input drive level is detected, the 5215A trips and locks into standby mode and displays a fault indicator.

 

Features

  • Designed for use with 5700A, 5200A, 0r 5100B
  • Output voltages to 1100V rms
  • Maximum output power 220 watts
  • 10 Hz to 100 kHz, typical upper limit 120 kHz
  • 420 ppm midband amplitude uncertainty at 1000 rms
  • 200 ppm midband six-month stability
  • Fully programmable
  • Short and overload protected
Download Data Sheet
Print
Email

Specifications

Input Power

100V, 115V, 200V, 230Vac, ±10%, internal jumper selected, 50 Hz to 60 Hz, 1800 VA at full load.

Receptacle on rear panel for calibrator power.

Size

26.7 x 43.2 W x 62.7 cm (10.5 x 17 x 24.7 in)

Weight

54.5 kg (120 lb)