Description
If you want an instrument with remote programmability, consider the Fluke PM 6304 RCL meter. The PM 6304 has a basic accuracy of 0.1%, 3 test voltage levels and 204 test frequencies. The PM 6304C has an even higher basic accuracy, and can measure components or devices with an accuracy of 0.05%. Thanks to this high accuracy, the PM 6304C is ideally suited for calibrating the capacitance sourcing function of the high-performance Fluke Calibration 5500A Multi-Product Calibrator.
Features
- Select the test you need
- Easy-to-use, at-a-glance display
- Test frequencies from 50 Hz to 1 MHz (100 kHz for PM 6304)
- 0.1% basic accuracy (0.05% for PM 6304C)
- RS-232 or IEEE-488 interface
- AC test levels from 50 mV to 2V rms
- Internal or external bias
- DC test measurement (optional)
- 9 front panel setups
- Actual component test voltage/current readback
- Automatic zero trimming
- Contact check and deviation mode (PM 6306 only)
- Front-panel test posts for immediate 4-wire measurements
Specifications
|
AC Test Mode |
|||||
|
Test frequency |
|
||||
|
Test frequency accuracy |
|
||||
|
Test signal levels |
|
||||
|
Basic measurement accuracy at normal measurement mode |
|
||||
|
DC bias |
|||||
|
Internal |
|
||||
|
External |
|
||||
|
DC test mode (Optional) |
|||||
|
Test signal levels |
|
||||
|
Basic measurement accuracy |
|
||||
|
Contact check (PM 6306 only) |
|
|
Pass |
3 3Ω |
|
Fail |
3Ω (with indication of failed connection lead) |
|
Maximum measuring ranges |
|
|
Impedance/Resistance AC Z or RAC |
0.0000O to 250 MΩ |
|
Resistance DC R DC |
0.0000O to 50 MΩ |
|
Capacitance C |
0.00 pF to 31.8F |
|
Inductance L |
0.00 mH to 637 kH |
|
Quality factor Q |
0.000 to 1000 |
|
Dissipation factor D |
0.000 to 1000 |
|
Phase Angle F |
-179 to + 175 deg |
|
Voltage monitor VX |
0.1 mV to 2.00V |
|
Current monitor IX |
0.005 mA to 10.0 mA |
|
Maximum resolution |
|
|
Impedance/Resistance AC Z or RAC |
0.1 mΩ |
|
Resistance DC R DC |
0.1 mΩ |
|
Capacitance C |
0.01 pF |
|
Inductance L |
0.01 uH |
|
Quality factor Q |
0.001 |
|
Dissipation factor D |
0.001 |
|
Phase Angle F |
0.1 deg |
|
Voltage monitor VX |
0.1 mV |
|
Current monitor IX |
0.001 mA |
|
Circuit diagram |
|
|
Display |
1 of 7 different equivalent circuit diagrams |
|
Auto Mode |
|
|
Read-out |
Dominant and secondary parameter |
|
Equivalent circuit diagram |
Parallel for R+C, Serial for R+L |
|
Manual mode |
|
|
Read-out |
Dominant and secondary parameter or Z, F, D, Q, VX, IX |
|
Equivalent circuit diagram |
Parallel or serial selectable |
|
Average function |
|||||
|
Function |
Exponential averaging in continuous mode |
||||
|
Levels |
|
||||
|
Deviation mode (PM 6306 only) |
|
|
Relative range in respect to reference value |
-100% to +100% |
|
Measuring modes (Normal) |
|||||
|
Continuous |
2 measurements/s |
||||
|
Single |
Triggered via “TRIG” key |
||||
|
Test frequency |
|
||||
|
Read-out |
Display or via IEEE-488 or RS-232 interface |
||||
|
Measuring modes (Fast) |
|||||||
|
Max. speed |
10 measurements/s |
||||||
|
Test frequency |
|
||||||
|
Single |
Triggered via handler interface |
||||||
|
Read-out |
Via IEEE or RS-232 interface (display blanked) |
||||||
|
Binning |
|
|
Standard bins |
9 |
|
Special bins |
Bin “0” and bin “fail” |
|
Bin programming via |
IEEE-488 interface |
|
|
RS-232 interface |
|
|
Bin programmer (PM 6304 only) |
|
Bin limit programming |
Absolute or relative |
|
Trim function |
|
|
Open circuit |
Open circuit compensation Z > 100 kW |
|
Short circuit |
Short circuit compensation Z < 10W |
|
Protection against charged capacitors |
|||||||||
|
C < 2 mF |
|
||||||||
|
2 mF ≤ C ≤ 2 mF |
|
||||||||
|
Stored settings (non-volatile memory) |
|
|
Front panel settings |
9 + 1 (trim figures included) |
|
Bin settings |
9 + 1 |
|
Print measurements results |
|
|
Print measurements results |
Via RS-232 interface for serial printers |
|
Calibration interval |
|
|
Calibration interval |
1 year |
|
Environmental Conditions |
|||||
|
Operating Temperature |
0° to 50° C |
||||
|
Storage Temperature |
-40° C to 70° C |
||||
|
Power, battery life |
100/125/225/240V + 10% |
||||
|
Line frequency |
50/60 Hz
|
||||
|
Power Consumption |
|
||||
|
Safety |
According to CE-regulation 73/23 EN61010-1 CAT II, Pollution Degree 2 |
||||
|
EMC |
According to CE regulation 89/336: Emission according to EN 55011, Group 1 Class B, respectively CISPR 11. Immunity according to EN 50082-1, inclusive IEC 751-2, -3, -4 |
||||
|
Warm up Time |
|
||||
|
Dimensions and weight |
|||||
|
W x H x D |
315 x 105 x 405 mm (12.4 in x 4.13 in x 15.9 in) |
||||
|
Weight |
|
||||