NoiseKen ESS-L1611 – GT30R Basic Model ESD Simulator 16kV
Description
Cost-oriented basic model ESD simulator with a lightweight discharge gun attached, attachable. The output voltage can be selected either with 30kV (B3011) or maximum 16kV (L1611) and compliant to both EN/IEC61000-4-2 Standard (bothB3011 and L1611) and ISO10605 Standard (B3011).
Features
- EN/IEC 61000-4-2 Ed2 Compliant (B3011/L1611)
- ISO 10605 Ed2 Compliant (B3011/L1611) (Note: Optional gun head and CR units are required.)
- “3 pre-checking function” taking the confirmable test into the account (B3011/L1611)
- “Discharge detecting function” to realize the air-discharge confirmation (B3011/L1611)
- “Lightest Discharge Gun in the market “to lighten the continuous operation (GT-30R)
- “White LED Irradiator” to facilitate the visualization of the discharging point (GT-30R)
Specifications
ESD Simulator Main Unit Model: ESS-B3011/L1611
Items | ESS-B3011 specifications | ESS-L1611 specifications |
Polarity | Positive or Negative | |
Output voltage | 0.2 to 30.0kV +/-5%
(30.5kV maximum) 0.2 to 10kV: 0.05kV step 10.0 to 30.5kV: 0.10kV step |
0.2 to 16.0kV +/-5%
0.2 to 10kV: 0.05kV step 10.0 to 16.0kV: 0.10kV step |
Repetition cycle | 0.05 to 99.9s (+/- 10%) /Manual
0.05 to 10.0s: 0.05s step 10.0 to 99.9s: 0.10s step |
|
No. of time of discharge | 1 to 999 times /continuous | |
Discharge mode | Contact discharge/ Air discharge | |
Trigger mode | Gun trigger / Main unit trigger | |
Operation panel | Indicator: 5×7 dot matrix LED
Operation: Push-buttons (Partially lighting) |
|
Radiation mode select switch | Extra / Normal switching function built-in | |
Discharge detection | Discharge detection function in air-discharge mode | |
Pre-checking function | High voltage output checking
(by user operation. Not the calibration but just checking) |
|
CR & Gun head checking | CR constant and gun head recognizable (with an indicator to prevent the wrong combination) | |
IEC level switching function | Contact discharge mode: 2.0, 4.0, 6.0, 8.0 kV step
Air discharge mode: 2.0, 4.0, 8.0, 15.0 kV step |
|
Warning lamp | Lights up while the HV circuitry is on.
Blinks while discharges are in progress. |
|
Charging capacitor and resistance | 150pF+/-10%, 330 ohm +/-10% (built into Discharge Gun GT-30R) | |
Main unit charging resistance | 10M ohm
(53Mohm for combination with Discharge Gun GT-30R) |
|
Power supply/ Power consumption | 100 to 240V AC 50/60 Hz +/-10% 62VA | |
Included accessories | Instruction manual | |
Dimensions and weight | (W)270 X (H)263 X (D)200 mm (projection excluded) Approx. 4.6 kg |
Discharge Gun Model: GT-30R
Items | GT-30R specifications |
Output voltage | 30.5kV Max |
Polarity | Positive or Negative |
Repetition cycle | 0.05s and over Note) Guaranteed value is 0.1s and over (330pF and under) |
Discharge mode | Contact or Air |
Charging resistor | Built-in CR unit |
Charge/Discharge capacitor & Discharge resistor | Built-in CR unit |
High voltage connector | Composite connector of high voltage and signal line (male) NoiseKen custom) |
Operation switch | Discharge trigger |
LED irradiator | White LED light built-in for irradiating EUT (Focusing point is the tip) |
Included accessories | CR unit (150pf-330 ohm) 06-00073A
Conical discharge tip 12-00007A Round discharge tip 12-00008A Instruction manual |
Dimensions and weight | (W)83 X (H)217 X (D)229 mm (projection excluded) Approx. 0.8kg (Cable and connector excluded)Approx. 10.8kg (Cable included, connector excluded) |