Temptronic ATS-850-M Thermostream -90° to +300°C
Description
Designed for 60Hz operation only, this Advanced Temperature Source for fast and precise thermal conditioning of components, parts, hybrids, modules, sub-assemblies, and printed circuit boards. Capable of ultra-low temperatures without the use of Liquid Nitrogen (LN2) or Liquid Carbon Dioxide (LCO2).
Features
- Transition Rates:
-55°C to +125°C, <10 seconds
+125°C to -55°C, <10 seconds - System Airflow Output: 4 to 18 scfm ( 1.8 to 8.5 l/s) Continuous
- Temperature Display & Resolution: ±1°C
- Temperature Accuracy: ±1°C
- Directly Control the device under test (DUT)
- DUT Control Sensors: Type T or Type K Thermocouples, RTD (100 Ohm)
- Remote interfaces: IEEE.488, RS232
- Thermal Cap provides a localized thermal test environment
protects from moisture and condensation at the DUT site - LabView™ drivers
- CE Compliant and CFC-free
- Heated Defrost to quickly remove moisture buildup from the internal chiller
- Heat Only Mode to reduce power usage when cold temperatures are not used
- Automatic Power Reduction to reduce power usage during idle periods
Specifications
Facility Requirements | ||
Power | 60Hz only, system does not operate at 50Hz
200 – 250 VAC (230V nominal), 60Hz, 30 amp, 1phase |
|
Compressed Air | ||
Clean, Dry Air (CDA) | Filtered to 5 micron particulate contamination.
Oil Content: <0.1 ppm, by weight, filtered to 0.01 micron oil contaminant. Dew point: <-10°C @ 6.2 BAR (90PSI) |
|
Air Supply Pressure | 6.2 to 7.6 BAR (90 to 110 PSIG) | |
Total Air Flow Rate Required | 7.1 to 14.2 l/s (15-30 scfm)
11.8 l/s (25 scfm) nominal |
|
Air Supply Temperature | +20° to +25°C; +22°C nominal | |
Operating Environment | ||
Operating Temperature | +20° to +28°C; +23°C nominal | |
Humidity | 0 to 60%; 45% nominal | |
Service and Safety | ||
Refrigerants | HCFC and CFC-free, non-toxic, non-flammable | |
Serviceability | Auto-diagnostics and field replaceable modules | |
Over Temperature Protection | +230°C (factory set): Operator can set high and low air temperature limits | |
Weight and Dimensions | ||
Dimensions | Base | Width: 61.0 cm (24 in.), Depth: 72.4 cm (28.5 in.)
Height: 108 cm (42.5 in.) |
System Weight | Not packed: 236 kg (520 lbs.)
Packed: 365 kg (805 lbs.) |
|
Mobility | Four static dissipative, swivel caster wheels | |
Max. Reach | 160.0cm (63 in.) | |
Max. Operating Height | 130.3 cm (51.2 in.) – extended height option: 188.0 (74.0 in.) | |
Min. Operating Height | 69.1 cm (27.2 in.) – extended height option: 81.3 (32.0 in.) | |
Noise Level | <65dBA |