Temptronic TPO4310A Thermal Inducing System -80°C to +225°C

Description

The Thermal Inducing Systems are used in a variety of temperature testing and conditioning applications. From their traditional applications in semiconductor testing, failure analysis, and device characterization to their broader use in PCB and electronic sub-assembly testing, the THERMOSTREAM has a system with the temperature capabilities to meet your needs. The THERMOSTREAM Systems offer an innovative temperature testing solution that allows you to perform device, board, and module testing right where you need it – at your test bench, in your production facility, or in your lab.

 

Features

  • Modes of Operation Two – Manual and Program Modes
  • Test set-up configurations – 12 Manual Mode test programs and 12 Program Mode test programs may be created
  • Test set-up storage – 11 Manual Mode test programs, 11 Program Mode test programs and 11 affiliated “Define Parameters” test set-ups can be stored for later recall.
  • Ramp/soak/cycle configurations –  In Program Mode, up to 12 sequences per test set-up; table is displayed on-screen
  • DUT temperature control – Patented4 Dual Loop Temperature Control
  • Status indicators – Real-time status displayed on-screen and via remoteinterface
  • Purge flow tester interface – Dry air purge protects tester electronics protection from condensation
  • Purge flow capacity – 0.25 to 1.5 liters per second (0.5 to 3 scfm) airflow
  • Calibration – Automated, simplified and accurate for all airflows and DUT types
  • Thermal head raising and lowering – Pneumatic control on thermal head, operated manually or via remote interface
  • Head positioner movement – Manual locking (4 locks), 360° head rotation; head can be manually pivoted, turned, tilted and vertically swung for ease of interface at tester site.
  • Manipulator (arm) movement – Motorized raising and lowering of arm; 330° positioning range around the base unit
Print
Email

Specifications

Temperature Range 1,3 -80° to +225°C (60 Hz System)
-75° to +225°C (50 Hz System)
Typical Temperature Transition Rate (air)  -55° to +125°C: approximately 10 seconds or less2
+125° to -55°C: approximately 10 seconds or less2
System air flow output 2.4 to 9 liters/second (5 to 18 scfm) CONTINUOUS
Temperature accuracy 1.0°C (when calibrated against NIST transfer standard)
Temperature set, display and resolution ± 0.1°C
Temperature Control DUT Sensor Ports Type T and Type K Thermocouple ports standard
DUT Control Control to within ± 0.1°C; SELF-TUNING available in DUT Control
Remote interface ports IEEE-488, RS232C Serial, and Start Test/End of Test/Stop on First Fail (ST/ET/SFF)
Power Requirements 200-250 VAC (230V nominal)
50 Hz, 30 amp, 1 phase
200-250 VAC (230V nominal)
60 Hz, 30 amp, 1 phase
Weights and Dimensions Base: Width: 61.0 cm (24 in.) Depth: 71.1 cm (28 in.); Height: 108 cm (42.5 in.); System weight: 236 kg (520 lbs); Packed: 365 kg (805 lbs)