Tenney T2 Temperature Humidity Cycling Chamber -75°C to 200°C

Description

T2 Series – Wide Range, Small Footprint The Tenney T2 temperature and humidity cycling chamber simulates a wide range of temperature and humidity conditions while utilizing minimal floor space. We specifically designed these environmental chambers to have a compact exterior and an ample interior workspace. The T2 Temperature and Humidity Cycling Chamber design allocates space within the interior of the cabinet for all standard options. This greatly improves the mobility of the unit as well as the overall appearance.

 

Features

  • Vapor-tight, continuously welded stainless steel interiors
  • Low workspace height to ease the loading and unloading of product
  • Vertical recirculation conditioning system
  • EIA-232 and EIA-485 data communications with Modbus RTU protocol
  • Platinum Resistance Temperature Detector (RTD)
  • Electronic capacitive sensor
  • Chamber overtemperature protection
  • Cascade refrigeration system
  • Built in safety devices
  • Explosion relief

Performance

  • Temperature Range:  -75°C to +200°C
  • Humidity capability: 20% to 95% RH in the dry bulb range of +20°C to 85°C as limited by a 3°C dewpoint. Test data based on 24°C ambient, sea level, 60Hz.
  • On 50 Hz or higher than 24°C ambient, performance may be reduced. Consult factory regarding any special cooling requirements. CFC-free refrigerants are used exclusively on all Tenney chambers.

Controller – VersaTenn III Controller

  • Two Channel
  • Profiling: 10 Profiles, 99 Steps Max.
  • Automatic & Manual Control
  • 2 Process Inputs: 100 Ohm Platinum RTD Sensor for each chamber
  • 1 Event Input
  • 10 Temp. Control Outputs: Time Proportioned and On / Off Control
  • Alarm Output
  • Guaranteed Soak
Print
Email

Specifications

Model 

T2RC

T2C

Workspace

W

16(406)

16(406)

D

13(330)

13(330)

H

15(381)

15(381)

Exterior

W

26(660)

27(686)

D

26(660)

27(686)

H

67(1702)

65(1651)

 Temperature

T2RC

T2C

Low

C

-75

-75

High

C

+200

+200

Humidity capability

20% to 95% RH in the dry bulb range of +20°C to 85°C as limited by a 3°C dewpoint.

Note: Test data based on 24°C ambient, sea level, 60Hz. On 50 Hz or higher than 24°C ambient, performance may be reduced. Consult factory regarding any special cooling requirements. CFC-free refrigerants are used exclusively on all Tenney chambers.

Change Rates in Minutes

Ambient to C Chamber Empty

200°C

90 Minutes

150°C

40 Minutes

100°C

25 Minutes

93°C

17 Minutes

65°C

9 Minutes

2°C

5 Minutes

-7°C

7 Minutes

-12°C

10 Minutes

-40°C

30 Minutes

-54°C

45 Minutes

-65°C

70 Minutes

-73°C

100 Minutes

-75°C

Ultimate

  Utilities, etc

Refrigeration

(2) 1/2) HP+ 1/2 HP

(2) 1/2) HP + 1/2 HP

Heater Capacity

500W

500W

Humidifier

 

W

500

GPH

0.15

AMPS @ 115V. 1Ph

18

18

AMPS Fuse

20

20

Unit Weight

LBS/KG

350/158

350/158