Tenney TC Series Cycling Chamber -73°C to 200°C
Description
Temperature Humidity Cycling Test Chambers Tenney's temperature and humidity cycling test chambers are capable of simulating a wide range of temperature or temperature and humidity conditions. The ten cubic foot work space will readily accept equipment as large as a 19"relay rack. In keeping with the needs of today's lab, Tenney designed these temperature humidity cycling test chambers to have a compact exterior that maximizes the interior workspace.
Features
- Watlow F4 controller
- Solid state humidity sensor
- Platinum RTD temperature sensor
- Heavy duty leveling casters
- Nichrome wire air heater
- Solid state pressure switches
- Backup safety thermal cutoff
- Chamber vent port
- Communications – RS485 and RS232
Controller – Watlow F4
- Single or Dual Channel
- Profiling: Forty Profiles, 256 Steps
- Automatic & Manual Control
- 1 Input: Universal Type (RTD is Std.)
- 4 Digital Inputs
- 2 Auxiliary Analog Inputs – Optional
- 2 Control Outputs: Time Proportioned, On/Off, or 4-20 ma
- 2 Alarm Outputs
- 8 Digital Event Outputs
- 2 Retransmit Outputs – Optional
- Communications: RS232 and RS485
Specifications
Model |
TC10S |
TC10C |
TC10RS |
TC10RC |
Cubic Feet |
10 |
|||
Liters |
283 |
|||
Work Space (W x D x H) |
24″ x 27.5″ x 28″ / 61cm x 69cm x 71cm |
|||
Exterior (nominal) (W x D x H) |
33.625″ x 59.121″ x 73.888″ / 86cm x 151cm x 188cm |
|||
Unit Weight (LBS) |
1000 |
1154 |
1000 |
1154 |
Temperature Range |
-35°C |
-73°C |
-35°C |
-73°C |
200°C |
200°C |
200°C |
200°C |
|
Humidity Range* |
– |
– |
20% – 95%RH |
20% – 95%RH |
Cooling Rate (empty chamber) Min (C/min) |
||||
180 to -65°C |
– |
77 (3.1) |
– |
77 (3.1) |
180 to -35°C |
71 (3) |
56 (3.8) |
71 (3) |
56 (3.8) |
125 to -35°C |
60 (2.6) |
41 (3.9) |
60 (2.6) |
41 (3.9) |
85 to -35°C |
53 (2.2) |
30 (4) |
53 (2.2) |
30 (4) |
85 to -20°C |
28 (3.7) |
25 (4.2) |
28 (3.7) |
25 (4.2) |
ambient to -35°C |
31 (1.9) |
16 (3.7) |
31 (1.9) |
16 (3.7) |
ambient to -68°C |
– |
38 (2.4) |
– |
38 (2.4) |
Heating Rate (empty chamber) Min (C/min) |
||||
-65 to 180°C |
– |
35 (7) |
– |
35 (7) |
-35 to 180°C |
30 (7.1) |
31 (6.9) |
30 (7.1) |
31 (6.9) |
-35 to 125°C |
20 (8) |
20 (8) |
20 (8) |
20 (8) |
-20 to 85°C |
11 (9.5) |
13 (8) |
11 (9.5) |
13 (8) |
-35 to 85°C |
10 (12) |
14 (8.5) |
10 (12) |
14 (8.5) |
-35°C to ambient |
6 (10) |
7 (8.5) |
6 (10) |
7 (8.5) |
-68°C to ambient |
– |
10 (9.3) |
– |
10 (9.3) |
Live Load Watts @ -18°C |
950 |
1450 |
950 |
1450 |
Live Load Watts @ -40°C |
– |
1050 |
– |
1050 |
Live Load Watts @ -54°C |
– |
800 |
– |
800 |
Temperature Control |
±0.3°C after stabilization |
|||
Temperature Uniformity |
±3°C @ 65°C |
|||
Humidity Control |
– |
– |
±2%RH |
±2%RH |
Air Flow Rate |
275 CFM |
|||
Heat Capacity |
3KW |
|||
Humidifier Capacity |
– |
– |
2KW |
2KW |
Refrigeration |
2HP |
(2) 2HP |
2HP |
(2) 2HP |
Condenser |
Air Cooled – Std Water Cooled Optional |
|||
Power |
208V – 230V 1P 60Hz |
|||
Amp – Fuse |
40 |
60 |
50 |
60 |
Amp – Under Load |
32 |
45 |
40 |
45 |
Power |
460V 1P 60Hz |
|||
Amp – Fuse |
20 |
30 |
30 |
30 |
Amp – Under Load |
16 |
23 |
20 |
23 |
Performance estimates based on 230v 1P 60Hz
*Humidity Range: based on temperature range 20°C to 85°C limited by 3°C dewpoint