Tenney Tenneystrat Altitude Test Chamber -70°C to 177°C
Description
Tenney Strat Altitude Chambers Tenney’s Tenneystrat Altitude Chambers simulate altitudes up to 200,000 feet. With an overall temperature range of -70°C to +177°C, humidity capabilities of 20% to 98% RH, and workspace sizes ranging from 5 to 64 cubic feet, this highly developed line of altitude chambers offers one of the broadest selections in the industry.
Features
- Vertical-down recirculating conditioning stream
- Highly efficient cascade refrigeration system with two semi-hermetic compressors
- Open-air nichrome wire heater elements
- Capacitance type absolute pressure transducer for pressure measurement
- Model T5STR- Non-corroding, vapor-injection humidity system with low-water protection
- Air pressure alarm output
Performance
- 70° C to + 177° C, ± 0.3° after stabilization
- Standard altitude conditions: Site level to 100,000 Feet (8.28 mm Hg / 8.28 Torr)
- Optional altitude conditions: Site level to 150,000 Feet (1.10 mm Hg / 1.10 Torr) Site level to 200,000 Feet (0.169 mm Hg / 0.169 Torr)
- Combined temperature and altitude: 60,000 feet maximum
- Standard humidity conditions (Models xxSTR only): 20% to 98% RH
- Control Accuracy: ±0.3°C and ±2% RH Controller VersaTenn V Controller
Controller
- Triple channel type controller
- Auto/ Manual Control
- Pentium class processor
- 8 MB Flash Disk On Chip storage
- USB Flash Disk port
- RS-232 communications
- Color touchscreen interface with real-time graphics, Windows CE Operating System, and latest software developments in test chamber control.
- Near infinite number of programs and steps
- 10 temperature control outputs
- 5 humidity control outputs
- 2 vacuum/ vent outputs
- 6 event outputs
Specifications
Model |
T5ST |
10ST |
15ST |
27ST |
36ST |
48ST |
64ST |
||||||||||||
System Horsepower Size |
4 |
6 |
15 |
6 |
15 |
6 |
15 |
20 |
6 |
15 |
20 |
15 |
20 |
30 |
15 |
20 |
30 |
||
Workspace |
W |
21/53 |
24/61 |
24/61 |
36/91 |
36/91 |
48/122 |
48/122 |
|||||||||||
D |
21/53 |
24/61 |
36/91 |
36/91 |
48/122 |
36/91 |
48/122 |
||||||||||||
H |
25/51 |
30/76 |
30/76 |
36/91 |
36/91 |
48/122 |
48/122 |
||||||||||||
Exterior |
W |
25/51 |
51/130 |
52/130 |
63/160 |
63/160 |
77/196 |
77/196 |
|||||||||||
D |
63/152 |
99/203 |
111/236 |
111/241 |
123/272 |
123-1/2 / 281 |
135-1/2 / 312 |
||||||||||||
H |
78/183 |
84/210 |
84/213 |
90/229 |
90/229 |
91/231 |
91/231 |
||||||||||||
Heater Capacity |
2 |
4 |
6 |
8 |
8 |
16 |
16 |
||||||||||||
Humdifier |
Watts |
1 |
2 |
3 |
|||||||||||||||
GPH |
1.05 |
1.0 |
1.5 |
||||||||||||||||
AMPS @ 230V, 3 |
33(1) |
48 |
81 |
52 |
85 |
63 |
96 |
130 |
63 |
96 |
130 |
120 |
149 |
183 |
120 |
149 |
183 |
||
AMPS, Fuse |
45 |
70 |
110 |
70 |
120 |
80 |
125 |
175 |
80 |
125 |
175 |
150 |
200 |
250 |
150 |
200 |
250 |
||
Cooling Water, 20°C GPM |
– |
3 |
6 |
3 |
6 |
3 |
6 |
9 |
3 |
6 |
9 |
6 |
9 |
12 |
6 |
9 |
12 |