Teseq GTEM 250A-SAE Test Cell for Emissions and Immunity Testing, SAE Version

Description

Teseq GTEM 250A-SAE Test Cell for SAE 1752/3 IEC 62132-2 & IEC 61967-2 cell is a test site for efficiently performing both radiated immunity and emissions testing in a single, controllable and shielded environment.

Teseq offers with GTEM 250A-SAE a GTEM cell with excellent VSWR for improved testing in range above 1 GHz.

Features

  • Test cell with special opening to test integrated circuits on approx. 45 mm septum height
  • Meets IEC/EN 61000-4-20, SAE J1752/3, IEC 62132-2 and IEC 61967-2
  • For 100 Watts input power
  • Excellent VSWR up to 18 GHz
Download Data Sheet
Print
Email

Specifications

Max. septum height 250 mm
Septum height at marker position 217 mm
Dimension (LxWxH in m) 1.25 x 0.65 x 0.45
Weight 45 kg
Height H1 of cell corpus 0.345 m
Height H2 of supports 0.105 m
Door (LxH in m) 0.20 x 0.13
EUT size (max. dimension, LxWxH in m) 0.20 x 0.20 x 0.15
EUT dimension for uniform-area 0 to 6 dB (LxWxH) in m 0.083 x 0.083 x 0.083
RF-input connector N-type
Nominal impedance 50 Ω
Frequency range DC up to 20 GHz
Frequency range according IEC/EN 61000-4-20 30 to 1000 MHz
Return loss / VSWR ≥ 20 dB / <1.25:1
Max input power (without additional external air cooling, without any EUT waste heat) below 1 GHz 100 W for 15 min
above 1 GHz 100 W continuous
Operating temperature +5°C to +30°C
Temperature range for this specification +20°C to 28°C
Shield characteristic (without any filter) 10 MHz to 1 GHz 100 dB
1 to 18 GHz 90 dB
Dimensions of the opening for IC testing 80 mm x 80 mm
Septum height for IC testing approx. 45 mm