Rod-L M25 Hipot/Ground Continuity Tester
Description
The Rod-L M25 is a 25 amp rated Ground Continuity Test Set. It allows testing of the ground system of a device under test to the rated test current. This insures that the ground and bond system will protect the operator during a catastrophic failure. A ground bond test applies a higher ac current to the ground connection. This test simulates the fault which may occur within a product by applying 25 A or higher at 60 Hz to the mains ground connection and any exposed metal on the product. A ground bond test checks the integrity of the connection and its ability to sustain high current. Most standards require that the resistance from power line ground to the conductive surface on the product does not exceed 0.1.
Features
- Performs ground continuity tests in full compliance with BSI, VDE and IEC
- Test time is adjustable from 1 second to 90 seconds
- Measures AC resistance from Chassis of Device Under Test (DUT) to power-cord ground lead
- Measures 0 to 0.15 ohm
- Audible and visual alarms identify faulty ground connections
- Hipot interface enables interconnecting to any ROD-L Hipot Test Instrument to form an automatic test system
- AC power source supplies 25, 30m 35 Amps (user specified) at Å 2.5 volts
- The Ground Continuity Test Instrument can be interconnected with any ROD-L Hipot Test Instrument to operate as a fully automatic system. One connection for DUT, and one start button to press
- Front panel receptacle is rated to 8000 VAC
- Front panel receptacle prevents director operator exposure to high voltage during test
Specifications
Output Current Resistance Test Time |
10, 25,30, or 35 amperes, user specif ied |
Resistance |
0 to 0.15 Ohm, Accurate to ± 5%, F.S |
Test Time |
1 second to 90 seconds, user adjustable (5 to 7 seconds- standard factory setting) |
Input Power Required |
115/230 VAC, ± 10%, 44-66Hz, 360 watts maximum |
Dimensions |
16.75″ X 5.25″ X 13.25″ (43cm x 13cm x 34cm) |
Weight |
30 lbs (13.6 kg) |
Color |
Mint Grey/Black |