EM TEST VSS 500N15.1 Voltage Surge Simulator

Description

VSS 500N15.1 Voltage Surge Simulator with 0.5J Constant Pulse Energy up to 15kV

EM TEST VSS 500N15.1 Voltage Surge Simulator generates high voltage transients as required by IEC 60255-5 with a source impedance of 500ohm for transient overvoltage tests on “Protection Relays”. The output voltage ranges up to 15kV, divided into two voltage ranges (up to 6.6kV and up to 15kV).

Testing to IEC 60255-5 standard requires a fixed energy of 0.5J at each test level which corresponds to no-load waveshape per IEC 61180-1. The output of the VSS 500N15.1 is divided into two ranges; the lower range covers the test up to 6.6kV while the higher range supports testing between 6kV and 15kV.

Pre-programmed Standard Test routines conveniently allow the user to change parameters during the test. Fail inputs are offered as standard for DUT monitoring purposes.

 

MODEL OVERVIEW

VSS 500N15.3 – 15kV Voltage Surge Simulator with 0.5J Constant Pulse Energy as per IEC 60255-27

Features

  • Test pulse 1.2/50us up to 15kV as per EN/IEC 60255-5
  • 7 different test levels
  • Source impedance 500 ohm
  • Constant energy of 0.5J at each test level
  • Peak voltage measurement
  • Peak current measurement
  • Interlock
  • Warning lamp control
  • Manual operation
  • USB and GPIB interface
Download Data Sheet
Print
Email

Specifications

TECHNICAL DETAILS
VOLTAGE SURGE SIMULATOR, PULSE 1.2/50US AS PER EN/IEC 60255-5
TECHNICAL DATA
Voltage (o.c.) 250V – 15,000V ± 10%
Front time 1.2µs ± 30%
Time to half value 50µs ± 20%
Source impedance 500ohm
Energy 0.5J ± 0.05J at each test level
Test level 0.55kV, 0.9kV, 3.00kV, 5.0kV, 6.6kV, 9.3kV and 14kV
Pulse forming As per IEC 60255-5
Polarity Positive, negative or alternating
Event counter 1 – 30,000 or endless
TRIGGER
Trigger of events Automatic, manual, external
CRO trigger 5V trigger signal for oscilloscope
Synchronization 0° – 360°, resolution 1°
Repetition rate 1s – 999s, depending on the voltage
OUTPUT
Direct Via HV connector; Zi = 500ohm

To connect external test boxes or devices

Output 1 500V – 6,600V
Output 2 6,600 – 15,000V
INTERFACE
Serial interface USB
Parallel interface IEEE 488, addresses 1 to 30
GENERAL DATA
Dimensions 19″/6HU
Weight approx. 25kg
Supply voltage 115/230V +10/-15%
Fuses 2xT2AT (230V) or 2xT4AT (115V)
Stay informed