AE Techron 2110 Gradient Amplifier
Description
AE Techron 2110 Linear Amplifier and Power Supply is a high-power, DC-enabled amplifier that offers a wide bandwidth and exceptional control of drift and distortion. The 2110 functions as a voltage or current source, making it a good choice for use in the lab or classroom. Its linear design provides a very-low noise floor and fast current rise times. Because the 2110 has no ripple noise, no synching with the console is required.
Features
- Up to 110 Ap, 195 Vp
- Current-mode response: DC-5 kHz (compensation dependent); Voltage-mode response: DC-20 kHz at rated power.
- Noise Floor (when Blanking circuit is enabled): 5 µA or less
- Current Settling Time on ramp of 0A to ±50A or ±50A to 0A: 20 µs to within 1.0A or 1%; 35 µs to within 200 mA or 0.2%
Specifications
Performance | ||
Specification typical at 25°C ambient. Unless otherwise noted; testing was done in Current mode with a load = 500 μH +100 mΩ | ||
Peak Current Limit | 112 A | |
Gain (adjustable) | Voltage Mode | 20 to 0.2 V/V µV (0.95 mV) |
Current Mode | 5 to 0.2 A/V | |
Gain Linearity
(over input signal, from 0.2V to 5V) |
DC | 0.0125% |
AC | 0.030% | |
Output Offset (adjustable to zero) | Voltage Mode | Less than ±400 μV |
Current Mode | ±5 mA | |
Input Characteristics | Three-Terminal Barrier Block Connector | Balanced with ground; 20 kΩ differential |
BNC Connector | Unbalanced; 10 kΩ single ended | |
Max Input Voltage | ± 10 V balanced or unbalanced | |
Common Mode Rejection | –58 dB with 5 V input | |
Output Impedance | Current Mode (effective) | 2000 Ω |
Voltage Mode (typical) | 3 mOhm in Series with 2.23 μH | |
Load | Current Mode | 500 μH + 100 mΩ |
Adaptable Range | 5 μH to 2.5 H, 0.01 Ω to 20 Ω | |
Current Mode Response | –3 dB at 5 kHz (compensation dependent) | |
Current Settling Time | Ramp 0 A to ±50 A or ±50 to 0 A | 20 μs to within 1.0 A or 1% |
35 μs to within 200 mA, 0.2% | ||
Total Harmonic Distortion | Current Mode | Less than 0.1% |
Load | 500 μH + 100 mΩ | |
Noise Floor (when Blanking circuit is enabled) | 5 μA or less | |
DC Drift | Self-Heating Drift, 0 to ±60 A | |
5 mA/10 minutes maximum | ||
Noise Output | 10 Hz to 1 kHz | 0.2 mA |
1 kHz to 60 kHz | 0.05 mA | |
Ripple Noise Output | None | |
Slew Rate, Voltage Mode | 41 V/μs | |
Remote Control and Monitoring
(back-panel D connector) |
Current Monitor | ± 1 V / 5 A ± 1% |
Reporting | System Fault, Over Temp, Over Voltage, Over Load | |
Control | Force to Standby, Remove from Standby, Reset after a Fault | |
Amplifier Protection | Over Load/Distortion (IOC) | Shutdown or clipped output |
Current vs Time (ODEP) | Clipped output | |
Each heat sink temperature | Shutdown 105°C | |
Overvoltage Shutdown | 229 VAC / 440 VAC | |
Undervoltage Shutdown | 187 VAC / 360 VAC | |
Status Indicators (front panel) | LEDs indicate a status of Run, Ready or Standby, and Fault, Over Temp, Over Voltage, and Over Load conditions | |
Controls (front panel) | AC Mains Switch and Circuit Breaker | Dual-function power switch and circuit breaker; rating: 20A for 208 volts or 15A for 400 volts. Turn off and then back on to reset. |
Connectors (back panel) | Power Connection | NEMA-style locking receptacle; matching AC connector also included |
Signal Output | 4-position terminal barrier block (OUTPUT/COMMON/ SAMPLED COMMON/CHASSIS GROUND); resistor between SAMPLED COMMON and CHASSIS GROUND terminals is a 2.7-ohm, 2W, 5%, metal-oxide resistor | |
Signal Input | User-selectable unbalanced BNC or balanced Barrier Strip | |
Interlock I/O Connection: | 25-pin D connector provides for remote monitoring and control functions | |
Power Requirements | Three-phase, 208 VAC ±10%, 47-60 Hz, 20 Amp AC service; (400 VAC ±10%, 15 Amp service model available) | |
Thermal Requirements | Operating Temperature | +10°C to +50°C (+50°F to +122°F). Maximum output power de-rated above 30° C (86°F) |
Storage | –30°C to +85°C (–22°F to +185°F) | |
Humidity | 70% or less, non-condensing | |
Physical Characteristics | Dimensions | 19” L x 8.75” H x 22.8” D (48.3 cm L x 22.3 cm H x 57.9 cm D) |
Cooling | Forced air cooling from front to back through removable filters via four 100 CFM fans. No space is required between rack-mounted amplifiers. Air filters are removable from the rear via one fastener per side and may be eliminated if cabinet filtration is provided. | |
Airflow | 400CFM | |
Weight | 103 lbs (46.7 kg) | |
Shipping Weight | 108 lbs (49.0 kg) |