EM TEST HSC 4-8 Coupling Network
Description
HSC 4-8 Coupling/Decoupling Assembly for Unshielded and Shielded High-Speed Communication Lines up to 1Gbit/s
EM TEST HSC 4-8 Coupling Network is used to couple Surge onto shielded and unshielded high-speed communication lines with data rates up to 1,000Mbit/s. Additionally, the HSC 4-8 can be used as a coupling/decoupling network to apply Burst- as well as Ringwave pulses to shielded and unshielded lines.
The HSC 4-8 meets the requirements according to IEC 61000-4-5.
Features
- Coupling to unshielded and shielded lines with up to 4 pairs
- Data transfer up to 1,000 Mbit/s
- Surge test voltage up to 3 kV (1.2/50 us)
- Burst test voltage up to 4 kV (5/50 ns)
- Ringwave test voltage up to 4 kV (0.5 us/100 kHz)
- Residual voltage max. 50 V at 3 kV Surge
- Compatible to PoE and PoE+ requirements
- DUT/AE connection via RJ45 connectors
Specifications
SURGE TEST MODE | |
Coupling mode | shielded and unshielded cable with up to 4 pairs |
Test voltage | Max. 3 kV* (1.2/50us)
(* only with connected RJ45 connector; if left open the max. test voltage is limited to 1.5 kV) |
Residual voltage | Max. 50V @3 kV Surge, line – PE |
RINGWAVE TEST MODE | |
Coupling mode | shielded and unshielded cable with up to 4 pairs |
Test voltage | Max. 4 kV* (0.5 us/100 kHz)
(* only with connected RJ45 connector; if left open the max. test voltage is limited to 1.5 kV) |
Residual voltage
line – PE |
Max. 50V @2 kV Ringwave
Max. 50V @4 kV Ringwave |
BURST TEST MODE | |
Coupling mode | shielded cable with up to 4 pairs |
Test voltage | Max. 4 kV* (5/50 ns)
(* only with connected RJ45 connector; if left open the max. test voltage is limited to 1.5 kV) |
Residual voltage
line – PE |
Max. 40 V @2 kV Burst
Max. 60 V @4 kV Burst |
TELECOM LINE SPECIFICATIONS | |
Input/output | RJ45 female connectors for telecom lines |
Terminal layout | Pair 1: pins 1/2
Pair 2: pins 3/6 Pair 3: pins 4/5 Pair 4: pins 7/8 |
Data rate | up to 1,000 MBit/s |
Protection | ± 60 V between two pairs |
Operating voltage | max. 57 VDC |
Operating current | max. 1 A |
INPUTS | |
Surge/Ringwave shield | SHV connector |
Surge unshielded | SHV connector |
Ringwave unshielded | SHV connector |
Burst | SHV connector |
GENERAL DATA | |
Dimensions (LxWxH) | 270 mm x 110 mm x 140 mm |
Weight | 6.5 kg |
Temperature | 5°C – 40°C (operation) |
Rel. humidity | 10 % – 90 %, non-condensing |