Rohde & Schwarz FPL1007 Spectrum Analyzer 5 kHz – 7 GHz
Description
Rohde & Schwarz FPL1007 Spectrum Analyzer 5 kHz – 7 GHz makes measuring fast and simple. The R&S®FPL1007 supports many applications such as power measurements and the analysis of digital and analog modulated signals. The solid RF performance makes it the ideal instrument for use in the lab, production and service. Its light weight and battery operation capability lets you take it anywhere to perform measurements.
MODEL SERIES OVERVIEW
FPL1000-series Spectrum Analyzers 5 kHz to 7.5 GHz
Features
- Frequency range 5 kHz to 7.5 GHz
- SSB phase noise: typ. –108 dBc (1 Hz) at 10 kHz (1 GHz carrier) offset
- DANL with preamplifier: typ. –166 dBm from 10 MHz to 2 GHz
- Lightweight with small footprint
- 40 MHz analysis bandwidth (option)
- Battery or 12 V/24 V DC operation (option)
Specifications
Frequency | ||
R&S®FPL1007 | 5 kHz to 7.5 GHz | |
Frequency resolution | 0.01 Hz | |
Bandwidth | ||
Resolution bandwidth (–3 dB) | sweep filters | 100 kHz to 10 MHz in 1/2/3/5 sequence |
FFT filters | 1 Hz to 50 kHz in 1/2/3/5 sequence | |
Signal analysis bandwidth (equalized) | standard | nom. 10 MHz |
with R&S®FPL1-B40 option | nom. 40 MHz | |
Displayed average noise level (DANL) | ||
RF preamplifier off | 5 kHz ≤ f < 100 kHz | typ. –143 dBm |
100 kHz ≤ f < 5 MHz | < –140 dBm, typ. –143 dBm | |
5 MHz ≤ f < 3 GHz | < –149 dBm, typ. –152 dBm. | |
3 GHz ≤ f < 5 GHz | < –143 dBm, typ. –146 dBm | |
5 GHz ≤ f ≤ 7.5 GHz | < –140 dBm, typ. –143 dBm | |
RF preamplifier on (gain nom. 20 dB) | 3 MHz ≤ f < 10 MHz | < –155 dBm, typ. –158 dBm |
10 MHz ≤ f < 2 GHz | < –163 dBm, typ. –166 dBm | |
2 GHz ≤ f < 3 GHz | < –162 dBm, typ. –165 dBm | |
3 GHz ≤ f < 5 GHz | < –158 dBm, typ. –161 dBm | |
5 GHz ≤ f < 7 GHz | < –156 dBm, typ. –159 dBm | |
7 GHz ≤ f < 7.5 GHz | < –155 dBm, typ. –158 dBm | |
Attenuation and Gain | ||
Setting range of attenuator | standard | 0 dB to 45 dB, in 5 dB steps |
with R&S®FPL1-B25 option | 0 dB to 45 dB, in 1 dB steps | |
PRF preamplifier gain | with R&S®FPL1-B22 option | nom. 20 dB |
Intermodulation | ||
1 dB compression of input mixer (two-tone) | RF attenuation 0 dB, RF preamplifier = off | nom. +7 dBm |
Third-order intercept point (TOI) | RF attenuation 0 dB, level 2 × –20 dBm, ∆f > 5 × RBW or 10 kHz, whichever is larger, RF preamplifier = off | |
10 MHz ≤ fin < 300 MHz | > 13 dBm, typ. 16 dBm | |
300 MHz ≤ fin < 3 GHz | > 17 dBm, typ. 20 dBm | |
3 GHz ≤ fin ≤ 7.5 GHz | > 15 dBm, typ. 18 dBm | |
with R&S®FPL1-B22 option, RF attenuation 0 dB, level 2 × –40 dBm, ∆f > 5 × RBW or 10 kHz, whichever is larger, RF preamplifier = on | ||
5 MHz ≤ fin < 7,5 GHz | nom. 0 dBm | |
Phase noise | ||
SSB phase noise | frequency = 1000 MHz, carrier offset | |
100 Hz | nom. –88 dBc (1 Hz) | |
1 kHz | < –99 dBc (1 Hz) | |
10 kHz | < –105 dBc (1 Hz), typ. –108 dBc (1 Hz) | |
100 kHz | < –110 dBc (1 Hz), typ. –115 dBc (1 Hz) | |
1 MHz | < –130 dBc (1 Hz), typ. –135 dBc (1 Hz) | |
10 MHz | nom. –152 dBc (1 Hz) | |
Overall Amplitude Accuracy | ||
Total measurement uncertainty | signal level 0 dB to –50 dB below reference level, S/N > 20 dB, sweep time auto, sweep type = FFT, RF attenuation 10 dB, 20 dB, 30 dB, 40 dB, RF preamplifier = off, span/RBW < 100, confidence level 95 %, +20 °C to +30 °C |
0.5 dB |
1 MHz ≤ f < 3 GHz | 0.5 dB | |
3 GHz ≤ f < 7.5 GHz | 0.8 dB |